Comparison of oxygenation measurements in pediatric patients during sickle cell crises
- PMID: 8283382
- DOI: 10.1016/s0022-3476(94)70260-8
Comparison of oxygenation measurements in pediatric patients during sickle cell crises
Abstract
Measurements of the saturation of arterial blood with oxygen (SaO2) were compared in 24 children during sickle cell crises. Simultaneous pulse oximetry (Nellcor N-100 pulse oximeter) and arterial blood analysis showed that SaO2 measured by pulse oximetry overestimated cooximeter-measured SaO2 (mean bias, 6.9%; p < 0.001). The blood gas machine-calculated SaO2 also overestimated cooximeter-measured SaO2 (p < 0.001). The bias increased with increasing age (p = 0.002) and carboxyhemoglobin level (p = 0.005) but was not related to methemoglobin, total hemoglobin, percentage of hemoglobin S, or percentage of hemoglobin F.
Publication types
MeSH terms
Substances