The performance of pinhole SPECT and the application of this technology to investigate the localization properties of radiopharmaceuticals in vivo in small laboratory animals are presented.
Methods: System sensitivity and spatial resolution measurements of a rotating scintillation camera system are made for a low-energy pinhole collimator equipped with 1.0-, 2.0- and 3.3-mm aperture pinhole inserts. The spatial detail offered by pinhole SPECT for in vivo imaging was investigated in studies of the brain and heart in Fisher 344 rats by administering 201TICI, 99mTc-HMPAO, 99mTc-DTPA and 99mTc-MIBI. Image acquisition is performed using a rotating scintillation camera equipped with a pinhole collimator; projection data are acquired in conventional step-and-shoot mode as the camera is rotated 360 degrees around the subject. Pinhole SPECT images are reconstructed using a modified cone-beam algorithm developed from a two-dimensional fanbeam filtered backprojection algorithm.
Results: The reconstructed transaxial resolution of 2.8 mm FWHM and system sensitivity of 0.086 c/s/kBq with the 2.0-mm pinhole collimator aperture provide excellent spatial detail and adequate sensitivity for imaging the regional uptake of the radiopharmaceuticals in tumor, organs and other tissues in small laboratory animals.
Conclusion: The resolution properties of pinhole SPECT are superior to those which have been achieved thus far with conventional SPECT or PET imaging technologies. Pinhole SPECT provides an important approach for investigating localization properties of radiopharmaceuticals in vivo.
Similar articles
-
High resolution pinhole SPECT for tumor imaging.Acta Oncol. 1993;32(7-8):861-7. doi: 10.3109/02841869309096147. Acta Oncol. 1993. PMID: 8305237
-
Ultra-high resolution SPECT system using four pinhole collimators for small animal studies.J Nucl Med. 1995 Dec;36(12):2282-7. J Nucl Med. 1995. PMID: 8523120
-
Evaluation of high-resolution pinhole SPECT using a small rotating animal.J Nucl Med. 2001 Dec;42(12):1863-9. J Nucl Med. 2001. PMID: 11752086
-
Ultra-high-resolution imaging of small animals: implications for preclinical and research studies.J Nucl Cardiol. 1999 May-Jun;6(3):332-44. doi: 10.1016/s1071-3581(99)90046-6. J Nucl Cardiol. 1999. PMID: 10385189 Review.
-
The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging.Eur J Nucl Med Mol Imaging. 2007 Feb;34(2):151-61. doi: 10.1007/s00259-006-0248-6. Eur J Nucl Med Mol Imaging. 2007. PMID: 17143647 Review.
Cited by
-
Analytic Determination of Rectangular-Pinhole Sensitivity With Penetration.IEEE Trans Med Imaging. 2020 Apr;39(4):833-843. doi: 10.1109/TMI.2019.2936187. Epub 2019 Aug 19. IEEE Trans Med Imaging. 2020. PMID: 31425068 Free PMC article.
-
Experimental evaluation of radiotracers: role of intact biological models.J Nucl Cardiol. 1998 Mar-Apr;5(2):167-83. doi: 10.1016/s1071-3581(98)90200-8. J Nucl Cardiol. 1998. PMID: 9588669 Review. No abstract available.
-
High resolution X-ray computed tomography: an emerging tool for small animal cancer research.Neoplasia. 2000 Jan-Apr;2(1-2):62-70. doi: 10.1038/sj.neo.7900069. Neoplasia. 2000. PMID: 10933069 Free PMC article. Review.
-
Use of tomographic nuclear medicine procedures, SPECT and pinhole SPECT, with cationic lipophilic radiotracers for the evaluation of axillary lymph node status in breast cancer patients.Eur J Nucl Med Mol Imaging. 2004 Jun;31 Suppl 1:S23-34. doi: 10.1007/s00259-004-1524-y. Epub 2004 Apr 15. Eur J Nucl Med Mol Imaging. 2004. PMID: 15085293 Review.
-
Myocardial infarction in rats: high-resolution single-photon emission tomographic imaging with a pinhole collimator.Eur J Nucl Med. 1996 Aug;23(8):896-900. doi: 10.1007/BF01084362. Eur J Nucl Med. 1996. PMID: 8753677
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources