[Use of bladder thermistor catheters in an intensive care unit. Comparative study of core temperature measurements with bladder thermometers and rectal thermometers in an intensive care unit]
- PMID: 8346579
[Use of bladder thermistor catheters in an intensive care unit. Comparative study of core temperature measurements with bladder thermometers and rectal thermometers in an intensive care unit]
Abstract
Bladder temperature measured by a thermistor tipped urinary catheter was compared to rectal temperature in 14 ICU patients. During a period of five days 2 x 232 measurements were recorded, and good correlation was found between rectal and bladder temperatures (R = 0.97). There did not seem to be any drift in the bladder thermistor. Measurement was accurate during periods of oliguria. We find that bladder temperature measurement is reliable, safe, convenient and accurate for routine use in the ICU.
Similar articles
-
Monitoring urinary bladder temperature in the intensive care unit: state of the science.Am J Crit Care. 2002 Jan;11(1):38-45; quiz 47. Am J Crit Care. 2002. PMID: 11785556 Review.
-
Urinary bladder and rectal temperature monitoring during clinical hypothermia.Nurs Res. 1989 Mar-Apr;38(2):73-6. Nurs Res. 1989. PMID: 2928149
-
A comparison of pulmonary artery, rectal, and tympanic membrane temperature measurement in the ICU.Heart Lung. 1993 Sep-Oct;22(5):435-41. Heart Lung. 1993. PMID: 8226008
-
Temporal artery versus bladder thermometry during perioperative and intensive care unit monitoring.Anesth Analg. 2007 Oct;105(4):1042-7, table of contents. doi: 10.1213/01.ane.0000281927.88935.e0. Anesth Analg. 2007. PMID: 17898385
-
[The choice of the method for body temperature measurement in intensive care patients: a literature review].Prof Inferm. 2010 Apr-Jun;63(2):99-105. Prof Inferm. 2010. PMID: 20943098 Review. Italian.