Implementation of the ETAR method for 3D inhomogeneity correction using FFT
- PMID: 8350813
- DOI: 10.1118/1.597010
Implementation of the ETAR method for 3D inhomogeneity correction using FFT
Abstract
The equivalent tissue-air-ratio (ETAR) method employs three-dimensional (3D) CT pixel information to approximate scatter dose contribution for inhomogeneity correction. In general, the calculation provides better agreement with measurements than the one-dimensional (1D) methods typically used in commercial treatment planning systems. In its original implementation, the 3D formulation of the ETAR method is modified empirically as a 2D calculation in order to reduce computation time. The modification compromises the use of the method in several treatment geometries. An examination of the ETAR formulation shows that the calculation can be expressed as a convolution and thus can be performed in 3D using fast Fourier transform (FFT) techniques. The algorithm has been implemented as a 3D FFT convolution. Making use of the symmetric properties of the FFT, the new approach shows significant savings in computation time without excessive memory requirement. Despite its fundamental limitations when applied to regions of electronic disequilibrium, the ETAR method offers a practical solution to improving current dose calculation in 3D treatment planning, particularly when the more advanced scatter ray-tracing dose calculation algorithms remain impractical for clinical use. Recent work to extend the method to approximate electron transport is also encouraging.
Similar articles
-
On methods of inhomogeneity corrections for photon transport.Med Phys. 1990 Sep-Oct;17(5):807-14. doi: 10.1118/1.596555. Med Phys. 1990. PMID: 2233566 Review.
-
Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine.Radiat Oncol. 2015 Apr 11;10:85. doi: 10.1186/s13014-015-0387-7. Radiat Oncol. 2015. PMID: 25885567 Free PMC article.
-
Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: current status and recommendations for clinical implementation.Med Phys. 2012 Oct;39(10):6208-36. doi: 10.1118/1.4747264. Med Phys. 2012. PMID: 23039658
-
Refinements of the finite-size pencil beam model of three-dimensional photon dose calculation.Med Phys. 1997 May;24(5):743-50. doi: 10.1118/1.597995. Med Phys. 1997. PMID: 9167166
-
A review of electron beam dose calculation algorithms.Australas Phys Eng Sci Med. 1996 Sep;19(3):111-30. Australas Phys Eng Sci Med. 1996. PMID: 8936722 Review.
Cited by
-
The effect of on-line position correction on the dose distribution in focal radiotherapy for bladder cancer.Radiat Oncol. 2009 Sep 24;4:38. doi: 10.1186/1748-717X-4-38. Radiat Oncol. 2009. PMID: 19775479 Free PMC article.
-
Dosimetric evaluation of a three-dimensional treatment planning system.J Med Phys. 2011 Jan;36(1):15-21. doi: 10.4103/0971-6203.75467. J Med Phys. 2011. PMID: 21430854 Free PMC article.
-
Computed tomography imaging parameters for inhomogeneity correction in radiation treatment planning.J Med Phys. 2016 Jan-Mar;41(1):3-11. doi: 10.4103/0971-6203.177277. J Med Phys. 2016. PMID: 27051164 Free PMC article. Review.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources