Patient dose and image quality in five standard x-ray examinations
- PMID: 8350841
- DOI: 10.1118/1.596987
Patient dose and image quality in five standard x-ray examinations
Abstract
Patient doses and image quality in five standard x-ray examination projections (22-36 measurements per each projection) have been measured, using homogeneous phantoms. The means and range of variation for the entrance skin air kerma values were: abdomen AP 5.2 mGy (1.1-21 mGy), chest PA 0.27 mGy (0.035-0.84 mGy), lumbar spine AP 8.4 mGy (2.9-28 mGy), skull LAT 3.1 mGy (1.1-7.7 mGy), and thoracic spine AP 7.9 mGy (1.3-22 mGy). Considerable variation was found in image quality parameters: film optical densities, image contrasts, and resolutions as well as in exposure values, tube voltages, exposure times, and in half-value thicknesses. The study indicated the need to standardize medical x-ray examination techniques in different x-ray departments. The measuring method can be used for comparing the patient dose and image quality of an individual x-ray unit to the doses and image quality in general use. The results can be used as the reference values for quality control of x-ray equipment.
Similar articles
-
PATIENT DOSES IN COMMON DIAGNOSTIC X-RAY EXAMINATIONS.Radiat Prot Dosimetry. 2019 Jul 1;184(1):12-27. doi: 10.1093/rpd/ncy169. Radiat Prot Dosimetry. 2019. PMID: 30289498
-
Entrance surface doses to patients undergoing selected diagnostic X-ray examinations in Sudan.Radiat Prot Dosimetry. 2007;123(2):209-14. doi: 10.1093/rpd/ncl137. Epub 2006 Sep 14. Radiat Prot Dosimetry. 2007. PMID: 16973669
-
Estimating pediatric entrance skin dose from digital radiography examination using DICOM metadata: A quality assurance tool.Med Phys. 2015 May;42(5):2489-97. doi: 10.1118/1.4918324. Med Phys. 2015. PMID: 25979042
-
Studying the dose level for different X-ray energy conventional radiography by TLD-100.Appl Radiat Isot. 2022 Mar;181:110066. doi: 10.1016/j.apradiso.2021.110066. Epub 2021 Dec 18. Appl Radiat Isot. 2022. PMID: 34968881 Review.
-
[CR (Computed Radiography)].Igaku Butsuri. 2002;22(2):110-7. Igaku Butsuri. 2002. PMID: 12766288 Review. Japanese.
Cited by
-
The use of reference image criteria in X-ray diagnostics: an application for the optimisation of lumbar spine radiographs.Eur Radiol. 2004 Sep;14(9):1561-7. doi: 10.1007/s00330-004-2320-6. Epub 2004 Apr 1. Eur Radiol. 2004. PMID: 15057564
-
Effective dose delivered by conventional radiology to Aosta Valley population between 2002 and 2009.Br J Radiol. 2012 Jul;85(1015):e330-8. doi: 10.1259/bjr/19099861. Epub 2011 Sep 21. Br J Radiol. 2012. PMID: 21937611 Free PMC article.
-
Local-Reference Patient Dose Evaluation in Conventional Radiography Examinations in Mazandaran, Iran.J Biomed Phys Eng. 2016 Jun 1;6(2):61-70. eCollection 2016 Jun. J Biomed Phys Eng. 2016. PMID: 27672626 Free PMC article.
-
Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations.J Appl Clin Med Phys. 2005 Winter;6(1):101-13. doi: 10.1120/jacmp.v6i1.2020. Epub 2005 Jan 12. J Appl Clin Med Phys. 2005. PMID: 15770201 Free PMC article.
-
New head equivalent phantom for task and image performance evaluation representative for neurovascular procedures occurring in the Circle of Willis.Proc SPIE Int Soc Opt Eng. 2012 Feb 23;8313:83130Q. doi: 10.1117/12.911351. Proc SPIE Int Soc Opt Eng. 2012. PMID: 24027618 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials