Percutaneous inferior vena caval filters: follow-up of seven designs in 320 patients
- PMID: 8351361
- DOI: 10.1148/radiology.188.3.8351361
Percutaneous inferior vena caval filters: follow-up of seven designs in 320 patients
Abstract
Three hundred twenty-four percutaneous inferior vena caval (IVC) filters of different designs were placed in 320 patients from April 1985 through June 1992. No acute mortality or substantial morbidity was attributed to filter placement. Radiologic or pathologic follow-up data were obtained in 227 (71%) patients (230 filters); clinical follow-up data only were obtained in 50 (16%) patients (50 filters). One hundred twenty (43%) patients died; post-filter-placement pulmonary emboli (PE) were related to the cause of death in eight. At IVC filter imaging studies, 26 of 137 (19%) filters demonstrated caval thrombus; 12 of 132 (9%) filters had delayed penetration through the IVC wall of greater than 3 mm; 13 of 230 (6%) filters migrated more than 1 cm; and five of 230 (2%) filters had fracture of a strut or leg. Deep venous thrombosis (DVT) at the insertion puncture site or in the lower extremity was noted in 26 of 117 (22%) cases of filter placement. Among patients without imaging studies, clinical suspicion of complications included PE in four patients, IVC thrombus in 14 patients, and lower-extremity DVT in 10 patients. Long-term clinical and radiologic follow-up of all IVC filters is indicated due to the relatively high prevalence of some complications.
Comment in
-
Complications of vena cava filters.Radiology. 1993 Sep;188(3):614-5. doi: 10.1148/radiology.188.3.8351319. Radiology. 1993. PMID: 8351319 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous