Echocardiographic determination of valid zero reference levels in supine and lateral positions
- PMID: 8353583
Echocardiographic determination of valid zero reference levels in supine and lateral positions
Abstract
Background: The phlebostatic axis--the junction of the fourth intercostal space and the midpoint of the anterior-posterior diameter--has been accepted as a reliable external reference point for the mid-right and mid-left atrium. Acceptance of this reference point is based upon research conducted in 1945 that measured venous pressures in the hands of subjects positioned with the head of the bed raised to different levels. The validity of this reference point for intracardiac pressure measurements in supine or laterally positioned patients has not been established.
Purpose: To determine the validity of the phlebostatic axis in the supine and lateral positions.
Methods: To determine validity in the supine position, we compared the distance from the phlebostatic axis to a fixed external point (the bed surface) and the distance from the right and left atria in the supine position to this same fixed external point. The distances from the right and left atria to the bed surface were determined with echocardiography and were used as the standard for the proper position of external reference points. To determine the validity of the phlebostatic axis in lateral positions, we compared the distances from the right atrium and left atrium to the bed surface in the supine position with those distances in different lateral positions.
Results: We analyzed the data of 25 normal, healthy subjects. The study findings show that the phlebostatic axis is a valid reference point for the right atrium, and the phlebostatic axis and midanterior-posterior diameter are valid reference points for the left atrium in the supine position. However, neither is a valid external reference point in the lateral positions. Pressure measurements obtained when patients are in the lateral positions are not accurate. There remains a need to develop valid methods of accurate pressure measurements in various body positions.
Similar articles
-
Reproducibility of central venous pressures in supine and lateral positions: a pilot evaluation of the phlebostatic axis in critically ill patients.Heart Lung. 1994 Jul-Aug;23(4):285-99. Heart Lung. 1994. PMID: 7960854 Review.
-
Reliability of measurements of pulmonary artery pressure obtained with patients in the 60 degrees lateral position.Am J Crit Care. 2000 Jan;9(1):43-51. Am J Crit Care. 2000. PMID: 10631390 Clinical Trial.
-
Comparisons of pulmonary artery pressure measurements in supine and 30 degree lateral positions.Can J Cardiovasc Nurs. 1995;6(3-4):4-8. Can J Cardiovasc Nurs. 1995. PMID: 8573278
-
Patient positioning and the accuracy of pulmonary artery pressure measurements (180f).Int J Nurs Stud. 1999 Dec;36(6):497-505. doi: 10.1016/s0020-7489(99)00033-4. Int J Nurs Stud. 1999. PMID: 10576120 Review.
-
Effect of position on selected hemodynamic parameters in postoperative cardiac surgery patients.Am J Crit Care. 1994 Jul;3(4):289-99. Am J Crit Care. 1994. PMID: 7920958 Clinical Trial.
Cited by
-
Evaluation of external reference levels for central venous pressure measurements of severely obese patients in the supine position.J Anesth. 2018 Aug;32(4):558-564. doi: 10.1007/s00540-018-2513-9. Epub 2018 May 28. J Anesth. 2018. PMID: 29808259
-
Optimal Transducer Level for Atrial and Pulmonary Arterial Pressure Measurement in Patients with Functional Single Ventricle.Pediatr Cardiol. 2017 Jan;38(1):44-49. doi: 10.1007/s00246-016-1481-9. Epub 2016 Oct 1. Pediatr Cardiol. 2017. PMID: 27696307
-
The jugular venous pressure revisited.Cleve Clin J Med. 2013 Oct;80(10):638-44. doi: 10.3949/ccjm.80a.13039. Cleve Clin J Med. 2013. PMID: 24085809 Free PMC article. Review.
-
[Central venous pressure. Validity, informative value and correct measurement].Anaesthesist. 2009 May;58(5):499-505. doi: 10.1007/s00101-009-1529-9. Anaesthesist. 2009. PMID: 19458975 Review. German.
-
Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position.J Clin Monit Comput. 2017 Apr;31(2):381-386. doi: 10.1007/s10877-016-9864-2. Epub 2016 Mar 21. J Clin Monit Comput. 2017. PMID: 27001384
Publication types