Insertional trauma of multichannel cochlear implants
- PMID: 8361322
- DOI: 10.1288/00005537-199309000-00010
Insertional trauma of multichannel cochlear implants
Abstract
Insertional trauma to the cochlea from three different multichannel cochlear implant electrodes was evaluated in a single-blind controlled study in fresh human temporal bones. Sixteen fresh human temporal bones were implanted with one of three types of multichannel electrodes (Symbion/InnerAid, Cochlear/Nucleus, or Storz/UCSF). Seven temporal bones were used as controls where a cochleostomy only was created. The temporal bones were evaluated histologically and cochlear histograms of the trauma were created. Although the three electrode designs caused damage which was unlikely to hinder implant performance, a distinct pattern of trauma was seen with each of the three electrode types. The least traumatic of the three electrode designs in this study was the Nucleus type. The degree of insertional trauma may be relevant to changing indications for insertion of cochlear implants as well as for patients with device failure who require reimplantation.
Similar articles
-
Impact of electrode insertion depth on intracochlear trauma.Otolaryngol Head Neck Surg. 2006 Sep;135(3):374-82. doi: 10.1016/j.otohns.2006.05.002. Otolaryngol Head Neck Surg. 2006. PMID: 16949967
-
Banded intracochlear electrode array: evaluation of insertion trauma in human temporal bones.Ann Otol Rhinol Laryngol. 1985 Jan-Feb;94(1 Pt 1):55-9. doi: 10.1177/000348948509400112. Ann Otol Rhinol Laryngol. 1985. PMID: 3838226
-
Combining perimodiolar electrode placement and atraumatic insertion properties in cochlear implantation -- fact or fantasy?Acta Otolaryngol. 2006 May;126(5):475-82. doi: 10.1080/00016480500437393. Acta Otolaryngol. 2006. PMID: 16698696
-
Cochleostomy site: implications for electrode placement and hearing preservation.Acta Otolaryngol. 2005 Aug;125(8):870-6. doi: 10.1080/00016480510031489. Acta Otolaryngol. 2005. PMID: 16158535 Review.
-
Cochlear implant and traumatic lesions secondary to electrode insertion.Rev Laryngol Otol Rhinol (Bord). 1998;119(5):317-22. Rev Laryngol Otol Rhinol (Bord). 1998. PMID: 10089801 Review.
Cited by
-
Micro-CT scan, electron microscopy and optical microscopy study of insertional traumas of cochlear implants.Surg Radiol Anat. 2015 Sep;37(7):815-23. doi: 10.1007/s00276-015-1469-9. Epub 2015 May 1. Surg Radiol Anat. 2015. PMID: 25929238
-
[Vestibular dysfunction after cochlear implant in adults].HNO. 2012 Oct;60(10):880-5. doi: 10.1007/s00106-012-2526-x. HNO. 2012. PMID: 22733278 Clinical Trial. German.
-
Morphometric Analysis of Temporal Bone Radiology for Cochlear Implant Candidacy.Indian J Otolaryngol Head Neck Surg. 2024 Feb;76(1):702-711. doi: 10.1007/s12070-023-04257-4. Epub 2023 Oct 29. Indian J Otolaryngol Head Neck Surg. 2024. PMID: 38440533 Free PMC article.
-
Modiolus-hugging intracochlear electrode array with shape memory alloy.Comput Math Methods Med. 2013;2013:250915. doi: 10.1155/2013/250915. Epub 2013 May 9. Comput Math Methods Med. 2013. PMID: 23762181 Free PMC article.
-
The effect of pulling out cochlear implant electrodes on inner ear microstructures: a temporal bone study.Int J Otolaryngol. 2011;2011:107176. doi: 10.1155/2011/107176. Epub 2011 Oct 11. Int J Otolaryngol. 2011. PMID: 22007225 Free PMC article.
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical