Controlled ventilation with the Mapleson D system A theoretical and experimental study
- PMID: 836743
- DOI: 10.1093/bja/49.2.107
Controlled ventilation with the Mapleson D system A theoretical and experimental study
Abstract
A theoretical analysis has been performed to show the interrelationships between alveolar carbon dioxide concentration, carbon dioxide output, alveolar ventilation and fresh gas flow during controlled ventilation with the Mapleson D system. The model and forms of equation used have been substantiated by experiments in which dogs were ventilated through a coaxial version of the Mapleson D system at varying levels of fresh gas flow and alveolar ventilation. By assuming that man ventilated with this system behaves as does the dog, a nomogram has been produced to predict alveolar carbon dioxide concentrations at any levels of fresh gas flow and minute volume.
Similar articles
-
Carbon dioxide distribution in Mapleson A and D systems: an experimental study.Acta Anaesthesiol Scand. 1989 Aug;33(6):439-43. doi: 10.1111/j.1399-6576.1989.tb02942.x. Acta Anaesthesiol Scand. 1989. PMID: 2508406
-
A theoretical study of gaseous homeostasis in the Magill circuit.Br J Anaesth. 1976 May;48(5):441-5. doi: 10.1093/bja/48.5.441. Br J Anaesth. 1976. PMID: 1276016
-
Preliminary evaluation of the Enclosed Magill Breathing System.Br J Anaesth. 1989 Feb;62(2):144-9. doi: 10.1093/bja/62.2.144. Br J Anaesth. 1989. PMID: 2493797
-
Mapleson's Breathing Systems.Indian J Anaesth. 2013 Sep;57(5):507-15. doi: 10.4103/0019-5049.120148. Indian J Anaesth. 2013. PMID: 24249884 Free PMC article. Review.
-
Electronic monitoring equipment.Semin Vet Med Surg Small Anim. 1993 May;8(2):119-26. Semin Vet Med Surg Small Anim. 1993. PMID: 8516584 Review. No abstract available.
Cited by
-
Carbon dioxide elimination during spontaneous ventilation with a modified Mapleson D system: studies in a lung model.Can Anaesth Soc J. 1978 Sep;25(5):353-65. doi: 10.1007/BF03006564. Can Anaesth Soc J. 1978. PMID: 279390 No abstract available.
-
A mass balance model for the Mapleson D anaesthesia breathing system.Can J Anaesth. 1993 Jun;40(6):554-67. doi: 10.1007/BF03009741. Can J Anaesth. 1993. PMID: 8403123
-
Capnography and the Bain circuit I: A computer model.J Clin Monit. 1985 Apr;1(2):103-13. doi: 10.1007/BF02832197. J Clin Monit. 1985. PMID: 3938477
-
Fresh gas flow in the Bain circuit during laparoscopy.Can J Anaesth. 1989 Jul;36(4):423-5. doi: 10.1007/BF03005342. Can J Anaesth. 1989. PMID: 2527090
-
The regulation of PaCO2 during controlled ventilation of children with a T-piece.Can Anaesth Soc J. 1979 Mar;26(2):104-13. doi: 10.1007/BF03013778. Can Anaesth Soc J. 1979. PMID: 288503 No abstract available.
MeSH terms
Substances
LinkOut - more resources