A comparison of the neurothesiometer and biothesiometer for measuring vibration perception in diabetic patients
- PMID: 8375265
- DOI: 10.1016/0168-8227(93)90006-q
A comparison of the neurothesiometer and biothesiometer for measuring vibration perception in diabetic patients
Abstract
The Neurothesiometer is a new device designed to replace the Biothesiometer in screening for diabetic peripheral neuropathy by measuring vibration perception threshold (VPT). It was compared with the Biothesiometer in 85 diabetic patients, mean age 61 (range 21-82) years and mean duration of diabetes 12 (1-26) years, attending routine diabetes and foot screening clinics. VPT correlated with age and duration of diabetes (r = 0.46 and r = 0.42, respectively P < 0.01), with both devices. The Biothesiometer VPT was 29.9 +/- 15.2 V (mean +/- S.D.) compared with 26.2 +/- 13.4 V with the Neurothesiometer (P = NS) and there was a close correlation between the two devices r = 0.93 P < 0.001. The within patient coefficient of variation was 8.6% for the Biothesiometer and 8.1% for the Neurothesiometer. The Neurothesiometer gives results that are in close agreement with the Biothesiometer and has a good coefficient of variation in routine clinical use. It is a self contained, battery operated device, which, although more expensive than its predecessor, is well suited to diabetes screening programmes and would be particularly useful to epidemiological surveys in the community.
Similar articles
-
A comparison of the new indicator test for sudomotor function (Neuropad) with the vibration perception threshold and the clinical examination in the diagnosis of peripheral neuropathy in subjects with type 2 diabetes.Exp Clin Endocrinol Diabetes. 2008 Feb;116(2):135-8. doi: 10.1055/s-2007-984455. Epub 2007 Dec 20. Exp Clin Endocrinol Diabetes. 2008. PMID: 18095233
-
Vibration perception threshold, complaints and sensory examination in diabetic patients.Acta Neurol Scand. 1991 May;83(5):297-300. doi: 10.1111/j.1600-0404.1991.tb04705.x. Acta Neurol Scand. 1991. PMID: 2063652
-
Assessment of the diagnostic accuracy of Vibrasense compared to a biothesiometer and nerve conduction study for screening diabetic peripheral neuropathy.J Foot Ankle Res. 2023 Sep 28;16(1):65. doi: 10.1186/s13047-023-00667-3. J Foot Ankle Res. 2023. PMID: 37770911 Free PMC article.
-
Comparison of vibration perception thresholds obtained with the Neurothesiometer and the CASE IV and relationship to nerve conduction studies.Diabet Med. 2002 Aug;19(8):661-6. doi: 10.1046/j.1464-5491.2002.00759.x. Diabet Med. 2002. PMID: 12147147 Clinical Trial.
-
Diagnostic Accuracy of Screening Tests for Diabetic Peripheral Neuropathy: An Umbrella Review.J Diabetes Res. 2024 Dec 4;2024:5902036. doi: 10.1155/jdr/5902036. eCollection 2024. J Diabetes Res. 2024. PMID: 39664106 Free PMC article. Review.
Cited by
-
Test-retest and time dependent variation and diagnostic values of vibratory sensation determined by biothesiometer and the Rydel-Seiffer tuning fork.Brain Behav. 2021 Aug;11(8):e2230. doi: 10.1002/brb3.2230. Epub 2021 Jun 4. Brain Behav. 2021. PMID: 34087955 Free PMC article.
-
Explanations for the lower rates of diabetic neuropathy in Indian Asians versus Europeans.Diabetes Care. 2010 Jun;33(6):1325-30. doi: 10.2337/dc09-2067. Epub 2010 Mar 9. Diabetes Care. 2010. PMID: 20215455 Free PMC article.
-
Medium-term outcomes of multi-disciplinary surgical management of non-ischemic diabetic heel ulcers.J Clin Orthop Trauma. 2021 Jan 30;17:30-36. doi: 10.1016/j.jcot.2021.01.012. eCollection 2021 Jun. J Clin Orthop Trauma. 2021. PMID: 33717969 Free PMC article.
-
Early retinal arteriolar changes and peripheral neuropathy in diabetes.Diabetes Care. 2012 May;35(5):1098-104. doi: 10.2337/dc11-1341. Epub 2012 Feb 28. Diabetes Care. 2012. PMID: 22374638 Free PMC article.
-
Fear of falling is prevalent in older adults with diabetes mellitus but is unrelated to level of neuropathy.J Am Podiatr Med Assoc. 2013 Nov-Dec;103(6):480-8. doi: 10.7547/1030480. J Am Podiatr Med Assoc. 2013. PMID: 24297984 Free PMC article.