Dependence on T1 of the echo amplitudes from multiple spin-echo sequences with equidistant echoes: simulation studies
- PMID: 8384283
- DOI: 10.1016/0730-725x(93)90024-8
Dependence on T1 of the echo amplitudes from multiple spin-echo sequences with equidistant echoes: simulation studies
Abstract
Transverse relaxation times were estimated from numerical simulations on spin systems using multi-echo spin-echo MRI protocols. The influence of T1 on the echo amplitudes via stimulated echo components was studied. The resulting effects on T2 estimates from the Carr-Purcell (CP), Carr-Purcell-Meiboom-Gill (CPMG), and Phase-Alternating-Phase-Shift (PHAPS; combination of CP and CPMG), multiple echo schemes were examined. Protocols with either spatially selective or nonselective refocusing pulses were studied. An intravoxel static field inhomogeneity of 0.1, 1, and 10 ppm was stimulated. The dependence on T1 of the T2 estimates was notable for T1 values below approximately 800 msec for all protocols. The PHAPS scheme provided rather accurate, but underestimated, T2 values when selective refocusing was used. With nonselective refocusing, PHAPS T2 values were overestimated and demonstrated a pronounced dependence on magnetic field inhomogeneity. In general, long T2 values were erroneous with the PHAPS protocol. The results indicate that a CPMG protocol structure provides a more robust method for T2 estimations than the PHAPS protocol.
Similar articles
-
Properties of the phase-alternating phase-shift (PHAPS) multiple spin-echo protocol in MRI: a study of the effects of imperfect RF pulses.Magn Reson Imaging. 1993;11(6):771-84. doi: 10.1016/0730-725x(93)90195-j. Magn Reson Imaging. 1993. PMID: 8371633
-
MRI measurements of the dependence on T1 of the echo amplitudes using a multiple spin-echo scheme.Magn Reson Imaging. 1993;11(8):1185-92. doi: 10.1016/0730-725x(93)90246-a. Magn Reson Imaging. 1993. PMID: 8271905
-
Approximate analytical expressions for the Carr-Purcell-Meiboom-Gill sequences: Decay rates and modulation zeros of the echo train and the relation between the T1 and T2 relaxation times.J Magn Reson. 2023 Jul;352:107464. doi: 10.1016/j.jmr.2023.107464. Epub 2023 May 2. J Magn Reson. 2023. PMID: 37148712
-
Magnetic Resonance Spectroscopy: Principles and Techniques: Lessons for Clinicians.J Clin Exp Hepatol. 2015 Dec;5(4):320-8. doi: 10.1016/j.jceh.2015.10.006. Epub 2015 Nov 12. J Clin Exp Hepatol. 2015. PMID: 26900274 Free PMC article. Review.
-
Recent 1D and 2D TD-NMR Pulse Sequences for Plant Science.Plants (Basel). 2021 Apr 21;10(5):833. doi: 10.3390/plants10050833. Plants (Basel). 2021. PMID: 33919424 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous