Studies on a new cross-axis coil planet centrifuge for performing counter-current chromatography. I. Design of the apparatus, retention of the stationary phase, and efficiency in the separation of proteins with polymer phase systems
- PMID: 8408404
- DOI: 10.1016/0021-9673(93)80703-b
Studies on a new cross-axis coil planet centrifuge for performing counter-current chromatography. I. Design of the apparatus, retention of the stationary phase, and efficiency in the separation of proteins with polymer phase systems
Abstract
An improved model of the cross-axis synchronous flow-through coil planet centrifuge has been designed in light of previous studies. The apparatus has a versatile feature in that both analytical and preparative columns can be accommodated in both off-center and central positions. Each has merit in separations. Retention of stationary phase was examined with various two-phase solvent systems used for the separation of biopolymers. Both analytical and preparative columns showed satisfactory retention of the stationary phase under optimum conditions. The apparatus was evaluated in separation of a set of protein samples using a polyethylene glycol-potassium phosphate biphasic system. In both types of columns all proteins were resolved with partition efficiencies of 260 to 670 theoretical plates. Further studies indicated that the relatively low partition efficiency of proteins is mainly attributed to their high molecular mass or molecular heterogeneity within each species rather than due to the high viscosity of the polymer phase system.
Similar articles
-
New angle rotor coil planet centrifuge for counter-current chromatography. II. Design of the apparatus and studies on phase retention and partition capability.J Chromatogr. 1986 May 23;358(2):325-36. doi: 10.1016/s0021-9673(01)90348-6. J Chromatogr. 1986. PMID: 3745355
-
Cross-axis synchronous flow-through coil planet centrifuge for large-scale preparative counter-current chromatography. I. Apparatus and studies on stationary phase retention in short coils.J Chromatogr. 1988 Sep 30;449(1):135-51. doi: 10.1016/s0021-9673(00)94374-7. J Chromatogr. 1988. PMID: 3235577
-
New small-scale cross-axis coil planet centrifuge. The design of the apparatus and its application to counter-current chromatographic separation of proteins with aqueous-aqueous polymer phase systems.J Chromatogr A. 2006 Feb 3;1104(1-2):245-55. doi: 10.1016/j.chroma.2005.12.046. Epub 2006 Jan 6. J Chromatogr A. 2006. PMID: 16405899
-
Recent advances in counter-current chromatography.J Chromatogr. 1991 Jan 18;538(1):3-25. doi: 10.1016/s0021-9673(01)91617-6. J Chromatogr. 1991. PMID: 2050794 Review.
-
Advances in countercurrent chromatography for protein separations.Bioanalysis. 2012 Apr;4(7):833-44. doi: 10.4155/bio.12.27. Bioanalysis. 2012. PMID: 22512800 Review.
Cited by
-
Partition Efficiency of High-Pitch Locular Multilayer Coil for Countercurrent Chromatographic Separation of Proteins Using Small-Scale Cross-Axis Coil Planet Centrifuge and Application to Purification of Various Collagenases with Aqueous-Aqueous Polymer Phase Systems.J Liq Chromatogr Relat Technol. 2011 Jan;34(3):182-194. doi: 10.1080/10826076.2011.546151. J Liq Chromatogr Relat Technol. 2011. PMID: 21869859 Free PMC article.
-
Alstonine as an antipsychotic: effects on brain amines and metabolic changes.Evid Based Complement Alternat Med. 2011;2011:418597. doi: 10.1093/ecam/nep002. Epub 2011 Jun 23. Evid Based Complement Alternat Med. 2011. PMID: 19189988 Free PMC article.
-
Partition Efficiency of Eccentric Coil for Countercurrent Chromatographic Separation of Proteins Using Small-scale Cross-axis Coil Planet Centrifuge with Circular and Elliptic Cylindrical Columns.Anal Sci. 2021 Feb 10;37(2):359-365. doi: 10.2116/analsci.20P292. Epub 2020 Oct 23. Anal Sci. 2021. PMID: 33100306
-
Purification of Proteins From Cell-Culture Medium or Cell-Lysate by High-Speed Counter-Current Chromatography Using Cross-Axis Coil Planet Centrifuge.Open Anal Chem J. 2007;1:28-37. doi: 10.2174/1874065000701010028. Open Anal Chem J. 2007. PMID: 25360182 Free PMC article.
-
PARTITION EFFICIENCY OF NEWLY DESIGNED LOCULAR MULTILAYER COIL FOR COUNTERCURRENT CHROMATOGRAPHIC SEPARATION OF PROTEINS USING SMALL-SCALE CROSS-AXIS COIL PLANET CENTRIFUGE WITH AQUEOUS-AQUEOUS POLYMER PHASE SYSTEMS.J Liq Chromatogr Relat Technol. 2009 Jan 1;32(8):1096-1106. doi: 10.1080/10826070902841547. J Liq Chromatogr Relat Technol. 2009. PMID: 20046931 Free PMC article.