Variability in measures of coronary lumen dimensions using quantitative coronary angiography
- PMID: 8409042
- DOI: 10.1016/0735-1097(93)90417-y
Variability in measures of coronary lumen dimensions using quantitative coronary angiography
Abstract
Objectives: The purpose of this study was to determine the true total variability of quantitative coronary angiographic measures and their components in the clinical setting.
Background: Many studies describe quantitative coronary angiographic variability on the basis of repeated quantitative coronary angiographic measures from the same cineangiogram. Although these studies characterize well the performance of quantitative coronary angiographic analysis methods, they do not include other potentially important sources of variability in results of this procedure, such as day to day variations in patients and equipment or variability in selection of frames for analysis.
Methods: Coronary angiograms from 20 patients who underwent diagnostic angiography followed by percutaneous transluminal coronary angioplasty an average of 2.9 days later were reviewed. A total of 30 lesions well visualized in both films were analyzed multiple times using an automated first-derivative edge-detection quantitative coronary angiographic technique.
Results: The coefficient of variation for quantitative coronary angiographic measures of the same lesions from separate angiograms ranged from 8.11% to 14.01%. Average diameter was the least variable and percent diameter stenosis the most variable. Day to day variations in the patient, procedure and equipment accounted for an average of 30% of the total variability. Of the remaining variability, only 13.26% was due to variability in frame selection.
Conclusions: These results provide useful information for planning clinical studies using quantitative coronary angiography, identify areas where additional improvements in this technology are needed and define more clearly the applicability of quantitative coronary angiography in the setting of routine clinical practice.
Similar articles
-
Variability of quantitative digital subtraction coronary angiography before and after percutaneous transluminal coronary angioplasty.Am J Cardiol. 1987 Jul 1;60(1):55-60. doi: 10.1016/0002-9149(87)90984-2. Am J Cardiol. 1987. PMID: 2955693
-
Immediate and late outcome of excimer laser and balloon coronary angioplasty: a quantitative angiographic comparison based on matched lesions.J Am Coll Cardiol. 1995 Oct;26(4):939-46. doi: 10.1016/0735-1097(95)00278-6. J Am Coll Cardiol. 1995. PMID: 7560621
-
Long-term luminal renarrowing after successful elective coronary angioplasty of total occlusions. A quantitative angiographic analysis.Circulation. 1995 Apr 15;91(8):2140-50. doi: 10.1161/01.cir.91.8.2140. Circulation. 1995. PMID: 7697842 Clinical Trial.
-
Definition and measurement of restenosis after successful coronary angioplasty: implications for clinical trials.Am Heart J. 1993 May;125(5 Pt 1):1394-408. doi: 10.1016/0002-8703(93)91013-5. Am Heart J. 1993. PMID: 8480594 Review.
-
Advantages and limitations of serial coronary arteriography for the assessment of progression and regression of coronary atherosclerosis. Implications for clinical trials.Circulation. 1993 Mar;87(3 Suppl):II38-47. Circulation. 1993. PMID: 8443922 Review.
Cited by
-
Lossy JPEG compression in quantitative angiography: the role of X-ray quantum noise.J Digit Imaging. 2011 Jun;24(3):516-27. doi: 10.1007/s10278-010-9275-8. J Digit Imaging. 2011. PMID: 20169463 Free PMC article.
-
Reproducibility of quantitative coronary analysis, Assessment of variability due to frame selection, different observers, and different cinefilmless laboratories.Int J Card Imaging. 1996 Sep;12(3):197-203. doi: 10.1007/BF01806223. Int J Card Imaging. 1996. PMID: 8915721
-
Impact of Cine Frame Selection on Quantitative Coronary Angiography Results.Clin Med Insights Cardiol. 2019 Mar 31;13:1179546819838232. doi: 10.1177/1179546819838232. eCollection 2019. Clin Med Insights Cardiol. 2019. PMID: 30967747 Free PMC article.
-
Restenosis after elective coronary balloon angioplasty in patients with end stage renal disease: a case-control study using quantitative coronary angiography.Heart. 1997 Oct;78(4):337-42. doi: 10.1136/hrt.78.4.337. Heart. 1997. PMID: 9404246 Free PMC article.
-
Visual and automatic grading of coronary artery stenoses with 64-slice CT angiography in reference to invasive angiography.Eur Radiol. 2007 Jun;17(6):1445-51. doi: 10.1007/s00330-006-0512-y. Epub 2006 Dec 16. Eur Radiol. 2007. PMID: 17180326
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources