MR microscopy of the rat lung using projection reconstruction
- PMID: 8419748
 - DOI: 10.1002/mrm.1910290117
 
MR microscopy of the rat lung using projection reconstruction
Abstract
Projection reconstruction has been implemented with self-refocused selection pulses on a small bore, 2.0 T MR microscope, to allow imaging of lung parenchyma. Scan synchronous ventilation and cardiac gating have been integrated with the sequence to minimize motion artifacts. A systematic survey of the pulse sequence parameters has been undertaken in conjunction with the biological gating parameters to optimize resolution and signal-to-noise (SNR). The resulting projection images with effective echo time of < 300 microseconds allow definition of lung parenchyma with an SNR improvement of approximately 15 x over a more conventional 2DFT short echo gradient sequence.