Evaluation of a one-piece poly(methyl methacrylate) intraocular lens with a 7 mm biconvex optic and a total diameter of 10 mm
- PMID: 8426314
- DOI: 10.1016/s0886-3350(13)80273-8
Evaluation of a one-piece poly(methyl methacrylate) intraocular lens with a 7 mm biconvex optic and a total diameter of 10 mm
Abstract
To find out how large the total diameter of an open-loop intraocular lens (IOL) should be, we studied 90 consecutive implantations of a one-piece poly(methyl methacrylate) IOL with a 7 mm biconvex optic and a total diameter of 10 mm (Adatomed 75 ST). The lens design made implantation simple, even through a small capsulorhexis. In 85% of the cases, the IOL remained well centered. One case with asymmetric placement of the IOL showed a decentration of more than 1 mm. In 14% of the eyes slight decentrations occurred because the capsular bag was too large. Contact between the iris and IOL was found in 15%. In 4% there were folds in the posterior capsule, and in 35% there was a small space between the IOL and the posterior capsule because the capsular bag was too large. Although the ST lens has a small total diameter, it remained at a distance from the iris in most cases. In-the-bag implantation through a continuous central capsulorhexis is mandatory to avoid secondary displacement; nevertheless, slight decentrations cannot be excluded. For a large capsular bag, the total diameter of an IOL should exceed 10 mm. Clinical results, cellular reaction on the lens surface (26.6%), biocompatibility, and visual performance are comparable to those of other well-approved poly(methyl methacrylate) IOLs.
Comment in
-
Blood in the capsular bag--endocapsular hematoma not a hyphema.J Cataract Refract Surg. 1993 Jul;19(4):566-7. doi: 10.1016/s0886-3350(13)80637-2. J Cataract Refract Surg. 1993. PMID: 8355176 No abstract available.
Similar articles
-
Experimental study of different intraocular lens designs implanted in the bag after capsulorhexis.J Cataract Refract Surg. 1996 Nov;22(9):1211-21. doi: 10.1016/s0886-3350(96)80070-8. J Cataract Refract Surg. 1996. PMID: 8972372
-
Decentration associated with asymmetric capsular shrinkage and intraocular lens design in a rabbit model.J Cataract Refract Surg. 1995 May;21(3):293-6. doi: 10.1016/s0886-3350(13)80135-6. J Cataract Refract Surg. 1995. PMID: 7674165
-
Capsulorhexis size and posterior capsule opacification.J Cataract Refract Surg. 1996 Jan-Feb;22(1):98-103. doi: 10.1016/s0886-3350(96)80277-x. J Cataract Refract Surg. 1996. PMID: 8656372
-
Centration of intraocular lenses with circular haptics.J Cataract Refract Surg. 1997 Oct;23(8):1247-53. doi: 10.1016/s0886-3350(97)80323-9. J Cataract Refract Surg. 1997. PMID: 9368172
-
Intraocular lenses.Curr Opin Ophthalmol. 1990 Feb;1(1):42-8. doi: 10.1097/00055735-199002000-00010. Curr Opin Ophthalmol. 1990. PMID: 10149666 Review. No abstract available.
Cited by
-
Bilateral identical intervals between phacoemulsification procedures performed 23 years before retinal detachment.Oman J Ophthalmol. 2018 Sep-Dec;11(3):274-276. doi: 10.4103/ojo.OJO_238_2017. Oman J Ophthalmol. 2018. PMID: 30505123 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources