. 1993 Jan-Feb;7(1):25-8.
doi: 10.1002/bmc.1130070107.

Measurement of terbutaline and salbutamol in plasma by high performance liquid chromatography with fluorescence detection

Affiliations

Measurement of terbutaline and salbutamol in plasma by high performance liquid chromatography with fluorescence detection

P T McCarthy et al. Biomed Chromatogr. 1993 Jan-Feb.

Abstract

A method is described for the determination of terbutaline and salbutamol in plasma from patients given maximal therapy for brittle asthma. The analytes were isolated by solid phase extraction on alkali-treated Bond-Elut, unmodified, silica columns and measured by high performance liquid chromatography with fluorescence detection (excitation wavelength 200 nm). The limits of detection for a 1 mL sample containing salbutamol and terbutaline were 1 microgram/L and 2.5 micrograms/L, respectively. The intra-assay precision (CV) for samples containing 25 micrograms/L was 3.6 and 5.0% respectively. This method was applied to the measurement of terbutaline in samples from patients given continuous infusions of the drug to assess whether this treatment might result in toxicity.