Comparative Study
. 1993 Feb;70(2):141-4.
doi: 10.1093/bja/70.2.141.

Comparison of automatic oscillometric arterial pressure measurement with conventional auscultatory measurement in the labour ward

Affiliations
Free article
Comparative Study

Comparison of automatic oscillometric arterial pressure measurement with conventional auscultatory measurement in the labour ward

M A Hasan et al. Br J Anaesth. 1993 Feb.
Free article

Abstract

We have compared two non-invasive methods of arterial pressure (AP) measurement used in labour wards: an automatic oscillometric measurement obtained by Dinamap 1846, and a conventional auscultatory measurement obtained by midwives. A total of 369 AP measurements were recorded, involving 28 normotensive and hypertensive pregnant women during labour, with or without extradural analgesia. Compared with the midwife group, the Dinamap group had a greater systolic AP, by 2.7 mm Hg (P < 0.01) and smaller diastolic AP, by 9.8 mm Hg (P < 0.01). The correlations between the two methods were highly significant, but the limits of agreement were relatively wide for both systolic and diastolic AP measurements. We conclude that a clinically important difference exists in diastolic AP measurements. Dinamap diastolic AP must be corrected using a regression equation, or simply by adding 10 mm Hg, before being compared with the available normal and hypertensive AP values.