Dosing efficiency and particle-size characteristics of pressurized metered-dose inhaler aerosols in narrow catheters
- PMID: 8449093
- DOI: 10.1378/chest.103.3.920
Dosing efficiency and particle-size characteristics of pressurized metered-dose inhaler aerosols in narrow catheters
Abstract
An experimental in vitro model was used to determine the effects of intraluminal catheter diameter and length on the delivered dose and particle-size characteristics of salbutamol (albuterol) aerosol delivered by metered dose inhaler (MDI) (Ventolin, 100 micrograms per puff). The dose of aerosolized drug that exited a 16-cm-long tracheal tube with an inner diameter (ID) of 6 mm was compared with that from 4 catheters of differing diameters and lengths that were inserted individually into the tracheal tube. The salbutamol MDI canister was actuated ten times into each delivery system, and the effluent aerosol was trapped onto a filter. The filtrate was dissolved in methanol, and the salbutamol concentration was determined using high-performance liquid chromatography. For the 3 22-cm-long catheters, the delivered dose (mean +/- SD) of salbutamol per actuation for the 22-standard wire gauge (SWG) catheter was 97.5 +/- 3.9 micrograms, which was similar to that for the 19-SWG catheter (102.3 +/- 2.5 micrograms) but was significantly less than that for the 14-SWG catheter (108.2 +/- 4.2 micrograms) (p < 0.05). These delivered doses exceeded those of the 6.0-mm-ID tracheal tube alone (2.33 +/- 0.76 micrograms) and the 13-cm-long 19-SWG catheter (2.17 +/- 0.29 micrograms) (p < 0.001). In a second experiment using a cascade impactor, the distribution of aerosol particle diameters that exited the 6-mm-ID tracheal tube was compared with that exiting a 13-cm-long 19-SWG catheter that extended halfway down the tracheal tube and with that exiting a 22-cm-long 19-SWG catheter inserted into the distal end of the 6-mm-ID tracheal tube. The mass median aerodynamic diameter (mean +/- SD) of the salbutamol aerosols delivered through both the 6.0-mm-ID tracheal tube (1.1 +/- 0.1 microns) and that of the 13-cm-long 19-SWG catheter (1.2 +/- 0.2 microns) were significantly less than that delivered through the 22-cm-long 19-SWG catheter (2.0 +/- 0.1 microns) (p < 0.05). The authors conclude that delivery of respirable aerosol can occur through narrow catheters that function as extended nozzles for MDIs. Optimal dosing will be obtained when the catheter extends the full length of the tracheal tube.
Comment in
-
Dosing efficiency and particle-size characteristics of pressurized metered-dose inhaler aerosols in narrow catheters.Chest. 1994 May;105(5):1621-2. doi: 10.1378/chest.105.5.1621b. Chest. 1994. PMID: 8181384 No abstract available.
-
Characteristics influencing the effective administration of drugs as inhalation aerosols.Chest. 1993 Mar;103(3):657-8. doi: 10.1378/chest.103.3.657. Chest. 1993. PMID: 8449044 No abstract available.
Similar articles
-
The delivery of aerosolized steroids from MDIs with nozzle extensions: quantitative laboratory evaluation of a method to improve aerosol delivery to intubated patients.Intensive Care Med. 1995 Nov;21(11):945-51. doi: 10.1007/BF01712338. Intensive Care Med. 1995. PMID: 8636529 Review.
-
High-efficiency delivery of salbutamol with a metered-dose inhaler in narrow tracheal tubes and catheters.Anesthesiology. 1991 Feb;74(2):360-3. doi: 10.1097/00000542-199102000-00024. Anesthesiology. 1991. PMID: 1990910
-
Combining a positive expiratory pressure device with a metered-dose inhaler reservoir system using chlorofluorocarbon albuterol and hydrofluoroalkane albuterol: effect on dose and particle size distributions.Respir Care. 2000 Mar;45(3):320-6. Respir Care. 2000. PMID: 10771801
-
Effects of Shake-Actuation Delay on Delivered Dose and Aerosol Characteristics of an Albuterol Suspension From a Pressurized Metered-Dose Inhaler.Respir Care. 2021 Aug;66(8):1315-1322. doi: 10.4187/respcare.08838. Epub 2021 Jun 8. Respir Care. 2021. PMID: 34103384
-
Effect of inhalation flow rate on the dosing characteristics of dry powder inhaler (DPI) and metered dose inhaler (MDI) products.J Aerosol Med. 1996 Summer;9(2):215-26. doi: 10.1089/jam.1996.9.215. J Aerosol Med. 1996. PMID: 10163351 Review.
Cited by
-
RNAi therapeutic strategies for acute respiratory distress syndrome.Transl Res. 2019 Dec;214:30-49. doi: 10.1016/j.trsl.2019.07.011. Epub 2019 Jul 27. Transl Res. 2019. PMID: 31401266 Free PMC article. Review.
-
The delivery of aerosolized steroids from MDIs with nozzle extensions: quantitative laboratory evaluation of a method to improve aerosol delivery to intubated patients.Intensive Care Med. 1995 Nov;21(11):945-51. doi: 10.1007/BF01712338. Intensive Care Med. 1995. PMID: 8636529 Review.
-
Preliminary study of electrocautery smoke particles produced in vitro and during laparoscopic procedures.Surg Endosc. 1996 Feb;10(2):152-8. doi: 10.1007/BF00188362. Surg Endosc. 1996. PMID: 8932618
-
VACuum INtubation (VACcIN) box restricts the exhaled air dispersion generated by simulated cough: description and simulation-based tests of an innovative aerosolization protective prototype.J Clin Monit Comput. 2022 Oct;36(5):1289-1295. doi: 10.1007/s10877-021-00759-0. Epub 2021 Sep 25. J Clin Monit Comput. 2022. PMID: 34562184 Free PMC article.
-
In vitro assessment of drug delivery through an endotracheal tube using a dry powder inhaler delivery system.Thorax. 1996 Jan;51(1):75-7. doi: 10.1136/thx.51.1.75. Thorax. 1996. PMID: 8658374 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources