Achieving accuracy for routine clinical chemistry methods by using patient specimen correlations to assign calibrator values. A means of managing matrix effects
- PMID: 8466405
Achieving accuracy for routine clinical chemistry methods by using patient specimen correlations to assign calibrator values. A means of managing matrix effects
Abstract
Accurate results obtained from routine methods used in the clinical laboratory can be achieved if the methods are traceable to definitive or reference methods or are based on widely accepted methodologic principles. At Eastman Kodak Co (Rochester, NY), we assign calibrator values for each method that is available on Kodak Ektachem analyzers by correlation to a reference method or a methodologic principle using patient specimens. Kodak's Reference Laboratory is responsible for maintaining the accuracy and consistency of the reference method or methodologic principle by standardizing with National Institute of Standards and Technology materials, where available, monitoring performance through statistical process control, participating in proficiency testing programs, and following procedures consistent with the International Organization for Standardization guideline, "General Requirements for the Competence of Calibrations and Testing Laboratories (ISO 25). "Patient samples that span the reportable range are run in replicate on the reference method or methodologic principle over several days. These patient sample results are then used to create a calibration curve based on the relationship of concentration to the raw response of the routine method, where the same patient samples are also run in replicate on the same day. Each day, multiple vials of stable calibrator fluids are included in these tests, along with control materials. Each calibrator is assigned a value of equivalent concentration (in millimoles per liter) by predicting the concentration that would be present at the measured response if the test fluid were a patient sample. These are termed supplementary assigned values (SAVs). The same calibrators can be used to compensate for matrix effects resulting from changes in reagent formulations, by appropriately adjusting the assigned values. For example, when a process improvement for Kodak Ektachem clinical chemistry slides for phosphorus was introduced, an SAV change of 0.39 mmol/L (1.2 mg/dL) was required for one calibrator to maintain accuracy of the patient samples. A comparison of patient samples, using the SAVs specific for each formulation, gave excellent correlation (in millimoles per liter; r2 = 0.995) by the following equation: New Formulation = 1.02 (Old Formulation) -0.016; Sy.x = 0.074. The benefits of this approach to calibration include (1) accuracy is traceable to a reference method or methodologic principle and, wherever available, to reference materials; (2) new calibrator values (SAVs) can be assigned whenever the reagent formulation changes to maintain accurate patient results; and (3) if matrix effects are present, they will not adversely influence accuracy because traceability is based on patient sample comparisons.
Similar articles
-
The accuracy of laboratory measurements in clinical chemistry: a study of 11 routine chemistry analytes in the College of American Pathologists Chemistry Survey with fresh frozen serum, definitive methods, and reference methods.Arch Pathol Lab Med. 1998 Jul;122(7):587-608. Arch Pathol Lab Med. 1998. PMID: 9674541
-
Matrix effects on proficiency testing materials. Impact on accuracy of cholesterol measurement in laboratories in the nation's largest hospital system.Arch Pathol Lab Med. 1993 Apr;117(4):345-51. Arch Pathol Lab Med. 1993. PMID: 8466396
-
Matrix effects and the accuracy of cholesterol analysis.Arch Pathol Lab Med. 1993 Apr;117(4):393-400. Arch Pathol Lab Med. 1993. PMID: 8466403
-
The Centers for Disease Control-National Heart, Lung and Blood Institute Lipid Standardization Program. An approach to accurate and precise lipid measurements.Clin Lab Med. 1989 Mar;9(1):105-35. Clin Lab Med. 1989. PMID: 2538292 Review.
-
The reference method value.Ann Ist Super Sanita. 1991;27(3):359-63. Ann Ist Super Sanita. 1991. PMID: 1809053 Review.
Cited by
-
Standardization in Clinical Enzymology.EJIFCC. 2009 Oct 12;20(3):141-7. eCollection 2009 Oct. EJIFCC. 2009. PMID: 27683339 Free PMC article.
-
Traceability of measurement results.Clin Biochem Rev. 2008 Aug;29 Suppl 1(Suppl 1):S61-6. Clin Biochem Rev. 2008. PMID: 18852860 Free PMC article.
-
Immunohistochemical demonstration of oestrogen and progesterone receptors: correlation of standards achieved on in house tumours with that achieved on external quality assessment material in over 150 laboratories from 26 countries.J Clin Pathol. 2000 Apr;53(4):292-301. doi: 10.1136/jcp.53.4.292. J Clin Pathol. 2000. PMID: 10823126 Free PMC article.
-
Interferences in clinical chemistry analysis.Indian J Clin Biochem. 1998 Jul;13(2):55-62. doi: 10.1007/BF02867865. Indian J Clin Biochem. 1998. PMID: 23105184 Free PMC article.