A model for technology assessment applied to pulse oximetry. The Technology Assessment Task Force of the Society of Critical Care Medicine
- PMID: 8472583
A model for technology assessment applied to pulse oximetry. The Technology Assessment Task Force of the Society of Critical Care Medicine
Abstract
Objectives: To test a model for the assessment of critical care technology. To develop practice guidelines for the use of pulse oximetry.
Data sources: A computer-assisted search of the English language literature and interviews with recognized experts in the field of pulse oximetry.
Study selection: Those studies that addressed one or more of the seven questions contained in our technology assessment template were analyzed. Study design was not a factor in article selection. However, the lack of well-designed clinical outcome studies was an important factor in determining the method of practice policy development we utilized.
Data extraction: A focus person summarized the data from the selected studies that related to each of the seven assessment questions. The preliminary data summary developed by the focus person was further analyzed and refined by the task force and then sent to 16 expert reviewers for comment. These expert comments were considered by the task force, and this final consensus report was developed.
Data synthesis: Pulse oximetry combines the principles of spectrophotometry and plethysmography to noninvasively measure oxygen saturation with a high degree of accuracy over the range of 80% to 100% saturation, assuming the device is being used according to the manufacturer's instructions and without any adverse operating conditions. The appropriate clinical uses of pulse oximetry fall into one of two broad categories: as a warning system based on continuous real-time measurement of arterial desaturation, or as an end-point for titration of therapeutic interventions. There are no published studies that allow for definitive, outcome-based conclusions concerning either the clinical impact or cost-benefit ratio of pulse oximetry.
Conclusions: The model developed for technology assessment proved to be appropriate for assessing pulse oximetry. The available data have allowed us to develop an evidence-based practice policy for the use of pulse oximetry in critical care. Critical care clinicians, researchers, and industry have a shared responsibility to provide valid outcome and efficacy studies of new technologies.
Comment in
-
Effectiveness and efficacy--which is which?Crit Care Med. 1994 Aug;22(8):1335-6. doi: 10.1097/00003246-199408000-00019. Crit Care Med. 1994. PMID: 8045154 No abstract available.
Similar articles
-
Guidelines for the transfer of critically ill patients. Guidelines Committee, American College of Critical Care Medicine, Society of Critical Care Medicine and the Transfer Guidelines Task Force.Am J Crit Care. 1993 May;2(3):189-95. Am J Crit Care. 1993. PMID: 8364669
-
A model for technology assessment as applied to closed loop infusion systems. Technology Assessment Task Force of the Society of Critical Care Medicine.Crit Care Med. 1995 Oct;23(10):1745-55. doi: 10.1097/00003246-199510000-00021. Crit Care Med. 1995. PMID: 7587242 Review.
-
Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health.CMAJ. 1992 Mar 1;146(5):703-12. CMAJ. 1992. PMID: 1562943 Free PMC article. Review.
-
Procedures and methods of benefit assessments for medicines in Germany.Eur J Health Econ. 2008 Nov;9 Suppl 1:5-29. doi: 10.1007/s10198-008-0122-5. Eur J Health Econ. 2008. PMID: 18987905
-
Next-generation pulse oximetry. Focusing on Masimo's signal extraction technology.Health Devices. 2000 Oct;29(10):349-70. Health Devices. 2000. PMID: 11094776
Cited by
-
Clinical impact of multi-parameter continuous non-invasive monitoring in hospital wards: a systematic review and meta-analysis.J R Soc Med. 2020 Jun;113(6):217-224. doi: 10.1177/0141076820925436. J R Soc Med. 2020. PMID: 32521195 Free PMC article.
-
Continuous intra-arterial blood gas monitoring.Intensive Care Med. 1994;20(2):85-6. doi: 10.1007/BF01707657. Intensive Care Med. 1994. PMID: 8201101 No abstract available.
-
Serial pulse oximetry in hepatopulmonary syndrome.Dig Dis Sci. 2011 Jun;56(6):1862-8. doi: 10.1007/s10620-011-1600-7. Epub 2011 Feb 16. Dig Dis Sci. 2011. PMID: 21327708
-
Technology assessment: a Canadian perspective.Can J Anaesth. 1996 May;43(5 Pt 2):R108-15. doi: 10.1007/BF03011673. Can J Anaesth. 1996. PMID: 8706213 Review. English, French. No abstract available.
-
Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output.Crit Care. 2004 Jun;8(3):190-5. doi: 10.1186/cc2852. Epub 2004 May 5. Crit Care. 2004. PMID: 15153237 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical