Dynamic field shaping for stereotactic radiosurgery: a modeling study
- PMID: 8478237
- DOI: 10.1016/0360-3016(93)90316-n
Dynamic field shaping for stereotactic radiosurgery: a modeling study
Abstract
Purpose: This work assesses the relative field shaping advantages of dynamic field shaping devices for stereotactic radiosurgery using a linear accelerator.
Methods and materials: We selected 43 intracranial tumors (2.0-4.2 cm maximum dimension, 1.5-25.5 cc tumor volume) out of the first 64 intracranial tumors treated with radiosurgery at the Joint Center for Radiation Therapy. We modeled five field shaping devices, each including a fixed auxiliary circular collimator: (a) fixed circular collimator alone; (b) two independent parallel jaws; (c) four independent rectangular jaws; (d) four independent rotatable jaws; and (e) "ideal" multileaf collimator. We adjusted the model parameters until the minimum target isodose was 80% of the dose delivered to isocenter. We defined the treatment volume ratio as the target volume divided by the treatment volume (volume receiving at least the minimum target dose). We used the treatment volume ratio to compare the five models and the actual patient treatments.
Results: For 34 tumors originally treated with one isocenter, the median Treatment Volume Ratio was higher for all of the device models except the fixed circular collimator compared to the actual patient treatments. For the nine tumors originally treated with multiple isocenters, the median Treatment Volume Ratio for the actual multiple isocenter treatments was similar to that for two parallel jaws, four rectangular jaws and four rotatable jaws. Only the median "ideal" collimator treatment volume ratio was higher for these nine tumors.
Conclusion: Simple field shaping devices have approximately 50% of the conformal advantage of an "ideal" multileaf collimator. Approximately 50% of typical radiosurgical tumors between 2 and 4 cm have field shaping advantages which exceed the geometrical uncertainties inherent in linear accelerator radiosurgery treatments. The three models, two parallel, four rectangular, or four rotatable independent jaws would improve current linear accelerator technology by providing homogeneous doses with equivalent field shaping for most tumors originally treated with inhomogeneous multiple isocenter plans (6/9 tumors in the current series).
Similar articles
-
Comparison of miniature multileaf collimation (MMLC) with circular collimation for stereotactic treatment.Int J Radiat Oncol Biol Phys. 1997 Feb 1;37(3):679-88. doi: 10.1016/s0360-3016(96)00507-x. Int J Radiat Oncol Biol Phys. 1997. PMID: 9112467
-
Impact of collimator leaf width on stereotactic radiosurgery and 3D conformal radiotherapy treatment plans.Int J Radiat Oncol Biol Phys. 1999 Jul 1;44(4):937-45. doi: 10.1016/s0360-3016(99)00041-3. Int J Radiat Oncol Biol Phys. 1999. PMID: 10386653
-
Dynamic arc radiosurgery field shaping: a comparison with static field conformal and noncoplanar circular arcs.Int J Radiat Oncol Biol Phys. 2001 Apr 1;49(5):1481-91. doi: 10.1016/s0360-3016(00)01537-6. Int J Radiat Oncol Biol Phys. 2001. PMID: 11286857
-
LINAC radiosurgery.Neurosurg Clin N Am. 1990 Oct;1(4):991-1008. Neurosurg Clin N Am. 1990. PMID: 2136180 Review.
-
M. Koecker on Diseases of the Jaws.Med Chir Rev. 1828 Apr 1;8(16):425-428. Med Chir Rev. 1828. PMID: 29917787 Free PMC article. Review. No abstract available.
Cited by
-
Quality of coverage: conformity measures for stereotactic radiosurgery.J Appl Clin Med Phys. 2003 Autumn;4(4):374-81. doi: 10.1120/jacmp.v4i4.2506. J Appl Clin Med Phys. 2003. PMID: 14604427 Free PMC article.
-
A novel correction factor based on extended volume to complement the conformity index.Br J Radiol. 2012 Aug;85(1016):e523-9. doi: 10.1259/bjr/27949149. Epub 2011 Nov 29. Br J Radiol. 2012. PMID: 22128127 Free PMC article.
-
Evaluation of a dedicated brain metastases treatment planning optimization for radiosurgery: a new treatment paradigm?Radiat Oncol. 2016 Feb 2;11:13. doi: 10.1186/s13014-016-0593-y. Radiat Oncol. 2016. PMID: 26831367 Free PMC article.
-
SPIDERplan: A tool to support decision-making in radiation therapy treatment plan assessment.Rep Pract Oncol Radiother. 2016 Nov-Dec;21(6):508-516. doi: 10.1016/j.rpor.2016.07.002. Epub 2016 Aug 24. Rep Pract Oncol Radiother. 2016. PMID: 27698591 Free PMC article.
-
Dosimetric performance of the new high-definition multileaf collimator for intracranial stereotactic radiosurgery.J Appl Clin Med Phys. 2010 Jun 21;11(3):3040. doi: 10.1120/jacmp.v11i3.3040. J Appl Clin Med Phys. 2010. PMID: 20717077 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical