Accuracy of a reflectance pulse oximeter in anesthetized horses
- PMID: 8484568
Accuracy of a reflectance pulse oximeter in anesthetized horses
Abstract
The accuracy of a reflectance pulse oximeter was determined in 22 anesthetized horses. Oximetric blood oxygen saturation was measured with the pulse oximeter probe attached to the mucosa of the mandible. Arterial blood oxygen saturation (SaO2) was calculated from arterial blood gas values and the equine blood oxygen dissociation curve. The mean +/- SD difference between oximetric blood oxygen saturation and SaO2 was -1.3 +/- 3.1% for values of SaO2 ranging from 80 to 100%. The difference between oximetric blood oxygen saturation and SaO2 was influenced by SaO2, the mean arterial blood pressure, and preanesthetic medication administered, but not by age, sex, or body weight of the horse, individual animal effect, anesthetic induction or maintenance agent, procedure performed, body position, mode of lung ventilation, time of sampling, arterial pH, or carbon dioxide tension.
Similar articles
-
Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation.Can J Vet Res. 2003 Jul;67(3):169-74. Can J Vet Res. 2003. PMID: 12889721 Free PMC article.
-
[Possibilities and limits of pulse oximetry for monitoring anesthesia in horses].Berl Munch Tierarztl Wochenschr. 1994 Jan;107(1):7-11. Berl Munch Tierarztl Wochenschr. 1994. PMID: 8129706 German.
-
Evaluation of pulse oximetry in anaesthetised foals using multiple combinations of transducer type and transducer attachment site.Equine Vet J. 1996 Nov;28(6):437-45. doi: 10.1111/j.2042-3306.1996.tb01615.x. Equine Vet J. 1996. PMID: 9049492
-
[Noninvasive patient observation in veterinary medicine: pulse oximetry and capnography. I. Pulse oximetry].Tierarztl Prax. 1994 Dec;22(6):596-606. Tierarztl Prax. 1994. PMID: 7716759 Review. German.
-
Limitations of pulse oximetry.Anesth Prog. 1992;39(6):194-6. Anesth Prog. 1992. PMID: 8250340 Free PMC article. Review.
Cited by
-
Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation.Can J Vet Res. 2003 Jul;67(3):169-74. Can J Vet Res. 2003. PMID: 12889721 Free PMC article.
-
Hypoxia driven opioid targeted automated device for overdose rescue.Sci Rep. 2021 Dec 31;11(1):24513. doi: 10.1038/s41598-021-04094-x. Sci Rep. 2021. PMID: 34972818 Free PMC article.
-
Comparison of Transmittance and Reflectance Pulse Oximetry in Anesthetized Dogs.Front Vet Sci. 2021 Apr 30;8:643966. doi: 10.3389/fvets.2021.643966. eCollection 2021. Front Vet Sci. 2021. PMID: 33996971 Free PMC article.
-
Assessment of clinical application of pulse oximetry probes in llamas and alpacas.Vet Med Sci. 2017 Jul 11;3(3):169-175. doi: 10.1002/vms3.68. eCollection 2017 Aug. Vet Med Sci. 2017. PMID: 29067213 Free PMC article.