An improved turbidimetric method for plasma fibrinogen
- PMID: 848496
An improved turbidimetric method for plasma fibrinogen
Abstract
Our experience in the use of Ellis' and Stransky's turbidimetric technique for the determination of plasma fibrinogen is presented. We standardized the test using a plasma pool whose fibrinogen content was determined by a weighed clot technique. The turbidimetric-fibrinogen test is accurate, sensitive, specific, reproducible, convenient, and inexpensive. Correlations between the turbidimetric method and the weighed clot reference method was shown. The assay is sensitive to as little as 25 mg/dl plasma fibrinogen, and the results are not affected by samples with a high content of lipids, bilirubin, fibrin split products, lymphangiogram dye, or therapeutic levels of heparin. Day-to-day replicate determinations on a plasma pool yielded a coefficient of variation of 3.4 percent. Results from as many as 10 plasma samples can be obtained in 30 minutes (10-minutes working time). Reagent cost is less than one cent per test.
Similar articles
-
The fibrinogen antigenic turbidimetric assay (FIATA): the X2x test--the corrected chi-square comparison against the control-mean.Clin Appl Thromb Hemost. 2007 Jan;13(1):73-100. doi: 10.1177/1076029606293436. Clin Appl Thromb Hemost. 2007. PMID: 17164498
-
A rapid turbidimetric determination of fibrinogen degradation products (FDP).Thromb Haemost. 1978 Feb 28;39(1):12-21. Thromb Haemost. 1978. PMID: 580487
-
A semiautomated turbidimetric method for determination of plasma fibrinogen.Am J Clin Pathol. 1978 Jul;70(1):76-8. doi: 10.1093/ajcp/70.1.76. Am J Clin Pathol. 1978. PMID: 696675
-
The fibrinogen functional turbidimetric assay.Clin Appl Thromb Hemost. 2008 Jan;14(1):84-96. doi: 10.1177/1076029607308031. Epub 2007 Dec 26. Clin Appl Thromb Hemost. 2008. PMID: 18160594
-
[Application of a bacterial endotoxin test for parenteral drugs].Eisei Shikenjo Hokoku. 1994;(112):209-11. Eisei Shikenjo Hokoku. 1994. PMID: 8854936 Review. Japanese.