Percutaneous thrombectomy of the acutely thrombosed dialysis graft: in vitro evaluation of four devices
- PMID: 8485746
- DOI: 10.1007/BF02602981
Percutaneous thrombectomy of the acutely thrombosed dialysis graft: in vitro evaluation of four devices
Abstract
Percutaneous recanalization of completely thrombosed hemodialysis grafts was simulated in a flow model. Thrombus removal was performed by 1) a Trac-Wright catheter; 2) an ultrasound-driven, oscillating probe aspiration thrombectomy (US-OAT) device; 3) an electric motor-driven oscillating probe aspiration thrombectomy (EM-OAT) device; and 4) a rotating spiral aspiration thrombectomy (RAT) device. All devices were able to restore a continuous lumen of the shunt model within 1-5 min, but residual layers of wall-adherent thrombi remained attached, particularly to the inner curvature of the shunt loop. Clot removal rate was 95% for the Trac-Wright catheter, 73% for the US-OAT device, 68% for the EM-OAT device (straight tip catheter), 88% for the EM-OAT device (angulated tip catheter), and 51% for the RAT device. Effluent particle rates for particles > 100 microns were 3.6% for the Trac-Wright catheter, 0.6% for US-OAT, and less than 0.1% for the other devices. Compared with the Trac-Wright catheter, the balance of clot-removing efficiency and creation of effluent particles favored the electric oscillating device with the angulated tip catheter, which had a slightly reduced, but still efficient, clot-removing capability and produced almost no effluent particles.
Similar articles
-
Angioscopic observations after percutaneous thrombectomy of thrombosed hemodialysis grafts.J Vasc Interv Radiol. 2000 Sep;11(8):971-7. doi: 10.1016/s1051-0443(07)61324-4. J Vasc Interv Radiol. 2000. PMID: 10997458
-
Mechanical Thrombectomy in Acute Thrombosis of Dialysis Arteriovenous Fistulae and Grafts Using a Vacuum-Assisted Thrombectomy Catheter: A Multicenter Study.J Vasc Interv Radiol. 2018 Jul;29(7):993-997. doi: 10.1016/j.jvir.2018.02.030. Epub 2018 May 30. J Vasc Interv Radiol. 2018. PMID: 29859664
-
Mechanical thrombectomy devices to treat thrombosed hemodialysis grafts.Tech Vasc Interv Radiol. 2003 Mar;6(1):35-41. doi: 10.1053/tvir.2003.36438. Tech Vasc Interv Radiol. 2003. PMID: 12772127 Review.
-
[Percutaneous mechanical thrombectomy with arrow-trerotola device inpatients with thrombosed graft fistula].Tani Girisim Radyol. 2003 Sep;9(3):371-6. Tani Girisim Radyol. 2003. PMID: 14661607 Turkish.
-
The use of the Fogarty catheter in 1998.Cardiovasc Surg. 1999 Apr;7(3):273-8. doi: 10.1016/s0967-2109(99)00007-1. Cardiovasc Surg. 1999. PMID: 10386742 Review.
Cited by
-
Evaluation of venous injury caused by a percutaneous mechanical thrombolytic device.Cardiovasc Intervent Radiol. 1995 May-Jun;18(3):172-8. doi: 10.1007/BF00204145. Cardiovasc Intervent Radiol. 1995. PMID: 7648594
-
Efficacy of a modified pharmacomechanical thrombolysis technique for endovascular treatment of thrombosed prosthetic arteriovenous grafts.Korean J Radiol. 2012 May-Jun;13(3):300-6. doi: 10.3348/kjr.2012.13.3.300. Epub 2012 Apr 17. Korean J Radiol. 2012. PMID: 22563267 Free PMC article.
-
New "mesh basket" for percutaneous removal of wall-adherent thrombi in dialysis shunts.Cardiovasc Intervent Radiol. 1993 Jan-Feb;16(1):7-10. doi: 10.1007/BF02603029. Cardiovasc Intervent Radiol. 1993. PMID: 8435840
References
MeSH terms
LinkOut - more resources
Medical