Percutaneous thrombectomy of the acutely thrombosed dialysis graft: in vitro evaluation of four devices

Affiliations

Percutaneous thrombectomy of the acutely thrombosed dialysis graft: in vitro evaluation of four devices

T Schmitz-Rode et al. Cardiovasc Intervent Radiol. 1993 Mar-Apr.

Abstract

Percutaneous recanalization of completely thrombosed hemodialysis grafts was simulated in a flow model. Thrombus removal was performed by 1) a Trac-Wright catheter; 2) an ultrasound-driven, oscillating probe aspiration thrombectomy (US-OAT) device; 3) an electric motor-driven oscillating probe aspiration thrombectomy (EM-OAT) device; and 4) a rotating spiral aspiration thrombectomy (RAT) device. All devices were able to restore a continuous lumen of the shunt model within 1-5 min, but residual layers of wall-adherent thrombi remained attached, particularly to the inner curvature of the shunt loop. Clot removal rate was 95% for the Trac-Wright catheter, 73% for the US-OAT device, 68% for the EM-OAT device (straight tip catheter), 88% for the EM-OAT device (angulated tip catheter), and 51% for the RAT device. Effluent particle rates for particles > 100 microns were 3.6% for the Trac-Wright catheter, 0.6% for US-OAT, and less than 0.1% for the other devices. Compared with the Trac-Wright catheter, the balance of clot-removing efficiency and creation of effluent particles favored the electric oscillating device with the angulated tip catheter, which had a slightly reduced, but still efficient, clot-removing capability and produced almost no effluent particles.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Dtsch Med Wochenschr. 1991 Jan 4;116(1):1-6 - PubMed
    1. AJR Am J Roentgenol. 1987 Jul;149(1):177-81 - PubMed
    1. Radiology. 1987 Sep;164(3):693-7 - PubMed
    1. Invest Radiol. 1991 Jun;26(6):557-63 - PubMed
    1. Cardiovasc Intervent Radiol. 1991 Mar-Apr;14(2):98-101 - PubMed

LinkOut - more resources