Prototype miniature endoluminal MR imaging catheter
Abstract
Purpose: The feasibility of a miniature endoluminal magnetic resonance (MR) detection coil was investigated for imaging mural and perimural anatomy of small, tubular structures.
Materials and methods: To this end, remotely tunable, single-loop, multiturn, receive-only radio-frequency coils, housed in 6-9-F arterial sheaths, were built. A 1.9-T imager was used. Phantom excitation was accomplished with a 62-mm-diameter bird-cage quadrature coil, and ex vivo specimen excitation was accomplished with a single-turn, untuned wire loop. Phantom images obtained with use of a 9-F catheter coil showed a signal-to-noise improvement on the border of 20 dB compared with images obtained with the quadrature coil. An 8-F catheter coil was used to obtain high-resolution (100 microns in-plane pixel size, 500 microns section thickness) spin-echo images (repetition time = 2,400 msec, echo time = 53 msec) of the wall of a fresh ex vivo human popliteal artery.
Results: Prospectively, these images were suggestive of the presence of diffuse intimal hyperplasia, medial calcification, and focal atherosclerotic plaque. These findings were confirmed histologically. Three-dimensional restacking of the axial images simplified examination of the normal layers and pathologic changes within the wall. The improved signal-to-noise characteristics of these miniature coils permit fast high-resolution imaging, allowing visualization of microscopic anatomic details.
Conclusions: With further development, this technology may be useful for studying atherosclerosis and for providing imaging guidance during endoluminal MR interventions.
Similar articles
-
High-resolution MR imaging appearance of colonic tissue in rabbits using an endoluminal coil.MAGMA. 2005 Nov;18(5):238-44. doi: 10.1007/s10334-005-0008-9. Epub 2005 Nov 18. MAGMA. 2005. PMID: 16320089
-
In vitro rat colonic wall imaging with MR endoluminal coil: feasibility study and histologic correlations.Acad Radiol. 2004 Jul;11(7):795-801. doi: 10.1016/j.acra.2004.01.022. Acad Radiol. 2004. PMID: 15217597
-
In vivo colon wall imaging using endoluminal coils: feasibility study on rabbits.J Magn Reson Imaging. 2004 Jul;20(1):90-6. doi: 10.1002/jmri.20059. J Magn Reson Imaging. 2004. PMID: 15221813
-
Intravascular MR imaging of atherosclerotic plaque: ex vivo analysis of human femoral arteries with histologic correlation.Radiology. 1997 Sep;204(3):769-74. doi: 10.1148/radiology.204.3.9280257. Radiology. 1997. PMID: 9280257
-
Radio-frequency coil selection for MR imaging of the brain and skull base.Radiology. 2001 Oct;221(1):11-25. doi: 10.1148/radiol.2211001537. Radiology. 2001. PMID: 11568316 Review.
Cited by
-
Design of catheter radio frequency coils using coaxial transmission line resonators for interventional neurovascular MR imaging.Quant Imaging Med Surg. 2017 Apr;7(2):187-194. doi: 10.21037/qims.2016.12.05. Quant Imaging Med Surg. 2017. PMID: 28516044 Free PMC article.
-
Magnetic resonance imaging-guided vascular interventions.Top Magn Reson Imaging. 2005 Oct;16(5):369-81. doi: 10.1097/00002142-200510000-00004. Top Magn Reson Imaging. 2005. PMID: 16924170 Free PMC article. Review.
-
High-resolution intravascular magnetic resonance quantification of atherosclerotic plaque at 3T.J Cardiovasc Magn Reson. 2012 Mar 26;14(1):20. doi: 10.1186/1532-429X-14-20. J Cardiovasc Magn Reson. 2012. PMID: 22448884 Free PMC article.
-
Real-time magnetic resonance imaging to guide pediatric endovascular procedures.Pediatr Cardiol. 2005 May-Jun;26(3):251-9. doi: 10.1007/s00246-005-1008-2. Pediatr Cardiol. 2005. PMID: 16132301
-
Microimaging of atherosclerotic plaque in animal models.MAGMA. 2000 Nov;11(1-2):58-60. doi: 10.1007/BF02678496. MAGMA. 2000. PMID: 11186989 No abstract available.
MeSH terms
LinkOut - more resources
Other Literature Sources
Medical
Miscellaneous