An assessment of a reflectance meter system for measurement of plasma or blood glucose in the clinic or side ward
- PMID: 852127
- DOI: 10.1016/0009-8981(77)90353-9
An assessment of a reflectance meter system for measurement of plasma or blood glucose in the clinic or side ward
Abstract
The Reflomat System for rapid estimation of plasma or blood glucose concentration has been evaluated. The System gave a linear response throughout its analytical range and the recovery of glucose added to glucose-free plasma was 97-105%. Addition of sodium fluoride to plasma produced a 7-15% reduction in the estimated glucose concentration. Plasma glucose concentration estimated with the Reflomat agreed closely with results of a glucose oxidase and a hexokinase based method, and blood glucose concentration measured with the Reflomat agreed well with results of a glucose oxidase method.