Comparison of PF4/heparin ELISA assay with the 14C-serotonin release assay in the diagnosis of heparin-induced thrombocytopenia
- PMID: 8526207
- DOI: 10.1093/ajcp/104.6.648
Comparison of PF4/heparin ELISA assay with the 14C-serotonin release assay in the diagnosis of heparin-induced thrombocytopenia
Abstract
The diagnosis of heparin-induced thrombocytopenia (HIT) may be affirmed by demonstrating heparin-dependent anti-platelet antibodies using the 14C-serotonin release assay (SRA). In this study, results of the SRA was compared with the recently described platelet factor 4 (PF4)/heparin enzyme-linked immunosorbent assay (ELISA). Compared with the SRA, the sensitivity and specificity of a PF4/heparin ELISA was 87% and 92%, respectively, using an assay developed in our laboratory; and 90% and 98%, respectively, using a commercially developed kit (Diagnostica Stago, Asnieres, France). However, antibodies to PF4/heparin were also detected in up to 8% of patients whose plasma was negative by SRA, and 23% of patients receiving heparin who were not thrombocytopenic. These data indicate that results obtained with the PF4/heparin ELISA and the SRA are generally in accord in patients with a clinical diagnosis of HIT. However, discrepant results occur in approximately 20% of cases because of the greater sensitivity of ELISA and the possible involvement of other heparin-binding proteins. The fact that each assay contributes independent information in some cases must be considered in the sequence of test performance and in providing consultation to the practicing hematologist.
Comment in
-
Diagnosis of heparin-induced thrombocytopenia. Still a journey, not yet a destination.Am J Clin Pathol. 1995 Dec;104(6):611-3. doi: 10.1093/ajcp/104.6.611. Am J Clin Pathol. 1995. PMID: 8526201 No abstract available.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous