Method specificity of non-invasive blood pressure measurement: oscillometry and finger pulse pressure vs acoustic methods
- PMID: 8554929
- PMCID: PMC1365146
- DOI: 10.1111/j.1365-2125.1995.tb04549.x
Method specificity of non-invasive blood pressure measurement: oscillometry and finger pulse pressure vs acoustic methods
Abstract
1. The agreement of blood pressure measurements by stethoscope auscultation (SBPa, DBPa-IV and DBPa-V), oscillometry (Dinamap; SBPo, and DBPo) and digital photoplethysmography (Finapres; SBPf, and DBPf) with the graphical analysis of the analogue microphone signals of vascular wall motion sound (SBPg and DBPg) was evaluated in eight healthy subjects in the presence of responses to the intravenous infusion of 1 microgram min-1 isoprenaline. 2. In general, there was good agreement between the SBP/DBP-measurements based on auscultatory Korotkoff-I- and IV-criteria and the reference method; the average method difference in estimating the isoprenaline responses for SBPa-SBPg was: -1.1, 95% CI: -5.4 to 3.1 mm Hg with a within-subject between-method repeatability coefficient (REP) of 11.6 mm Hg and for DBPa-IV-DBPg: 3.5, 95% CI: -0.5 to 6.5 mm Hg, REP: 11.5 mm Hg. The ausculatation of Korotkoff-V substantially overestimated the isoprenaline induced reduction of DBP: method difference DBPa-V-DBPg: -11.3, 95% CI: -17.8 to -4.7 mm Hg, REP: 31.8 mm Hg. 3. Oscillometry yielded good approximations for the SBP response to isoprenaline (average method difference SBPo-SBPg: -2.9, 95% CI: -9.0 to 3.3 mm Hg, REP: 17.6 mm Hg) but was poorly sensitive with regard to the DBP responses: method difference DBPo-DBPg: 6.5, 95% CI: -1.3 to 14.3 mm Hg, REP: 25.7 mm Hg. 4. Whilst the finger pulse pressure agreed well with regard to DBP (method difference for the DBP responses to isoprenaline: DBPf-DBPg: 1.8, 95% CI: -5.1 to 8.6 mm Hg, REP: 18.5 mm Hg) it was rather unsatisfactory with regard to SBP (method difference SBPf-SBPg: -14.1, 95% CI: -28.2 to -0.1 mm Hg, REP: 49.9 mm Hg).(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
Method specificity of the auscultatory estimates of the inodilatory reduction of diastolic blood pressure based on Korotkoff IV and V criteria.Br J Clin Pharmacol. 1995 May;39(5):485-90. doi: 10.1111/j.1365-2125.1995.tb04484.x. Br J Clin Pharmacol. 1995. PMID: 7669483 Free PMC article. Clinical Trial.
-
Variation of the Korotkoff Stethoscope Sounds During Blood Pressure Measurement: Analysis Using a Convolutional Neural Network.IEEE J Biomed Health Inform. 2017 Nov;21(6):1593-1598. doi: 10.1109/JBHI.2017.2703115. IEEE J Biomed Health Inform. 2017. PMID: 29136608
-
Does the position or contact pressure of the stethoscope make any difference to clinical blood pressure measurements: an observational study.Medicine (Baltimore). 2014 Dec;93(29):e301. doi: 10.1097/MD.0000000000000301. Medicine (Baltimore). 2014. PMID: 25546675 Free PMC article.
-
From Korotkoff and Marey to automatic non-invasive oscillometric blood pressure measurement: does easiness come with reliability?Expert Rev Med Devices. 2016;13(2):179-89. doi: 10.1586/17434440.2016.1128821. Epub 2016 Jan 8. Expert Rev Med Devices. 2016. PMID: 26641026 Review.
-
Blood pressure measurement in the hemodialysis setting.Adv Ren Replace Ther. 1997 Apr;4(2):174-8. doi: 10.1016/s1073-4449(97)70044-x. Adv Ren Replace Ther. 1997. PMID: 9113233 Review.
Cited by
-
Evaluation of noninvasive blood pressure recording by photoplethysmography in clinical studies using angiotensin challenges.Br J Clin Pharmacol. 1999 Oct;48(4):586-93. doi: 10.1046/j.1365-2125.1999.00049.x. Br J Clin Pharmacol. 1999. PMID: 10583030 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous