Five- to nine-year follow-up results of balloon angioplasty of native aortic coarctation in infants and children
- PMID: 8557921
- DOI: 10.1016/0735-1097(95)00479-3
Five- to nine-year follow-up results of balloon angioplasty of native aortic coarctation in infants and children
Abstract
Objectives: To evaluate the usefulness of balloon angioplasty for relief of native aortic coarctation, we reviewed our experience with this procedure, with special emphasis on follow-up results.
Background: Controversy exists with regard to the role of balloon angioplasty in the treatment of native aortic coarctation.
Methods: During an 8.7-year period ending September 1993, 67 neonates, infants and children underwent balloon angioplasty for native aortic coarctation. A retrospective review of this experience with emphasis on long-term follow-up forms the basis of this study.
Results: Balloon angioplasty produced a reduction in the peak-to-peak coarctation gradient from 46 +/- 17 (mean +/- SD) to 11 +/- 9 mm Hg (p < 0.001). No patient required immediate surgical intervention. At intermediate-term follow-up (14 +/- 11 months), catheterization (58 patients) and blood pressure (2 patients) data revealed a residual gradient of 16 +/- 15 mm Hg (p > 0.1). When individual results were scrutinized, 15 (25%) of 60 had recoarctation, defined as peak gradient > 20 mm Hg. Recoarctation was higher (p < 0.01) in neonates (5 [83%] of 6) and infants (7 [39%] of 18) than in children (3 [8%] of 36), respectively. Two infants in our early experience had surgical resection with excellent results. Three patients had no discrete narrowing but had normal arm blood pressure and had no intervention. The remaining 10 patients had repeat balloon angioplasty with reduction in peak gradient from 52 +/- 13 to 9 +/= 8 mm Hg (p < 0.001). Reexamination 31 +/- 18 months after repeat angioplasty revealed a residual gradient of 3 to 19 mm Hg (mean 11 +/- 6). Three (5%) of 58 patients who underwent follow-up angiography developed an aneurysm. Detailed evaluation of the femoral artery performed in 51 (88%) of 58 patients at follow-up catheterization revealed patency of the femoral artery in 44 (86%) of 51 patients. Femoral artery occlusion, complete in three (6%) and partial in four (8%), was observed, but all had excellent collateral flow. Blood pressure, echocardiography-Doppler ultrasound and repeat angiographic or magnetic resonance imaging data 5 to 9 years after angioplasty revealed no new aneurysms and minimal (2%) late recoarctation.
Conclusions: On the basis of these data, it is concluded that balloon angioplasty is safe and effective in the treatment of native aortic coarctation; significant incidence of recoarctation is seen in neonates and infants; repeat balloon angioplasty for recoarctation is feasible and effective; and the time has come to consider balloon angioplasty as a therapeutic procedure of choice for the treatment of native aortic coarctation.
Comment in
-
The various therapeutic approaches to aortic coarctation: is it fair to compare?J Am Coll Cardiol. 1996 Feb;27(2):471-2. doi: 10.1016/0735-1097(95)00547-1. J Am Coll Cardiol. 1996. PMID: 8557922 No abstract available.
Similar articles
-
Role of balloon angioplasty in the treatment of aortic coarctation.Ann Thorac Surg. 1991 Sep;52(3):621-31. doi: 10.1016/0003-4975(91)90961-o. Ann Thorac Surg. 1991. PMID: 1832851 Review.
-
[Balloon angioplasty for native coarctation in children: one year follow-up results].Zhonghua Er Ke Za Zhi. 2014 Jul;52(7):535-9. Zhonghua Er Ke Za Zhi. 2014. PMID: 25224061 Chinese.
-
Follow-up results of balloon angioplasty of native coarctation in neonates and infants.Am Heart J. 1990 Dec;120(6 Pt 1):1310-4. doi: 10.1016/0002-8703(90)90241-o. Am Heart J. 1990. PMID: 2147352
-
Immediate and follow-up results of balloon angioplasty of postoperative recoarctation in infants and children.Am Heart J. 1990 Dec;120(6 Pt 1):1315-20. doi: 10.1016/0002-8703(90)90242-p. Am Heart J. 1990. PMID: 2147353
-
Balloon angioplasty for coarctation of the aorta: immediate and long-term results.Am Heart J. 1988 Mar;115(3):657-65. doi: 10.1016/0002-8703(88)90817-4. Am Heart J. 1988. PMID: 2964188 Review.
Cited by
-
Balloon angioplasty for native aortic coarctation in different anatomic variants.Pediatr Cardiol. 2008 May;29(3):521-9. doi: 10.1007/s00246-007-9029-7. Pediatr Cardiol. 2008. PMID: 18043860
-
The role of stents in the treatment of congenital heart disease: Current status and future perspectives.Ann Pediatr Cardiol. 2009 Jan;2(1):3-23. doi: 10.4103/0974-2069.52802. Ann Pediatr Cardiol. 2009. PMID: 20300265 Free PMC article.
-
Intravascular stent therapy for coarctation of the aorta.Methodist Debakey Cardiovasc J. 2014 Apr-Jun;10(2):82-7. doi: 10.14797/mdcj-10-2-82. Methodist Debakey Cardiovasc J. 2014. PMID: 25114759 Free PMC article. Review.
-
Balloon Dilatation in the Management of Congenital Obstructive Lesions of the Heart: Review of Author's Experiences and Observations-Part I.J Cardiovasc Dev Dis. 2023 May 23;10(6):227. doi: 10.3390/jcdd10060227. J Cardiovasc Dev Dis. 2023. PMID: 37367392 Free PMC article. Review.
-
Long-term results (up to 18 years) of balloon angioplasty on systemic hypertension in adolescent and adult patients with coarctation of the aorta.Clin Cardiol. 2007 Feb;30(2):75-80. doi: 10.1002/clc.25. Clin Cardiol. 2007. PMID: 17326072 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources