Optimisation of transcutaneous cardiac pacing by three-dimensional finite element modelling of the human thorax

Affiliations

Optimisation of transcutaneous cardiac pacing by three-dimensional finite element modelling of the human thorax

D Panescu et al. Med Biol Eng Comput. 1995 Nov.

Abstract

The goal of the study is to determine by finite element analysis (FE) the optimal electrode placement, size and electrolyte resistivity that minimise the pain experienced by patients during successful transcutaneous cardiac pacing (TCP). The three-dimensional FE model generated for this purpose has 55,388 nodes, 50,913 hexahedral elements and simulated 16 different organs and tissues, as well as the properties of the electrolyte. The model uses a non-uniform mesh with an average spatial resolution of 0.8 cm in all three dimensions. To validate this model, the voltage across 3 cm2 Ag-AgCl electrodes is measured when currents of 5 mA at 50 kHz are injected into a subject's thorax through the same electrodes. For the same electrode placements and sizes and the same injected current, the FE analysis produced results in good agreement with the experimental data. The optimisation analysis tested seven different electrode placements, five different electrode sizes and six different electrolyte resistivities. The analysis indicates that the anterior-posterior electrode placement, electrode sizes of about 90 cm2 and electrolytes with resistivity of about 800 omega.cm yield the most uniform current distribution through the skin, thus having the best chances to minimise the pain delivered to the patient during successful TCP. The anterior-anterior electrode placement is the second most efficient.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Pacing Clin Electrophysiol. 1985 Nov;8(6):862-9 - PubMed
    1. Crit Care Med. 1981 May;9(5):393-4 - PubMed
    1. IEEE Trans Biomed Eng. 1987 Sep;34(9):743-8 - PubMed
    1. IEEE Trans Biomed Eng. 1993 May;40(5):434-9 - PubMed
    1. IEEE Trans Biomed Eng. 1993 Mar;40(3):246-55 - PubMed

Publication types