Evaluation of the Autobac 1 susceptibility testing system in a clinical diagnostic laboratory
- PMID: 856028
- PMCID: PMC352061
- DOI: 10.1128/AAC.11.4.750
Evaluation of the Autobac 1 susceptibility testing system in a clinical diagnostic laboratory
Abstract
The Autobac 1 system was compared to the disk diffusion and agar dilution methods. Overall, there was good correlation; however, certain antibiotic/bacterial combinations gave unreliable results. In particular, unreliable results were obtained against staphylococci with penicillins, aerobic gram-negative rods with cephalothin, and enterococci with cephalothin and clindamycin.