Infections during extended circulatory support: University of Alabama at Birmingham experience 1989 to 1994
- PMID: 8561607
- DOI: 10.1016/0003-4975(95)01021-1
Infections during extended circulatory support: University of Alabama at Birmingham experience 1989 to 1994
Abstract
Background: The University of Alabama at Birmingham experience with investigational ventricular assist devices (VADs) used as a bridge to transplantation has increased over the past several years; it now includes 27 VAD implantations with 13 VAD runs lasting for extended periods (ie, > 30 days). A review of complications experienced by patients during extended VAD runs is warranted before the further development and testing of chronically implanted mechanical circulatory support devices.
Methods: This study focuses on the infectious complications of extended VAD support; it includes 13 patients who were supported by either a Thoratec or HeartMate VAD for longer than 30 days pending cardiac transplantation. Infection was defined as any positive culture. The infections were classed according to site and severity as follows: class I were patient-related non-blood-borne infections, class II were blood-borne infections, class III were VAD percutaneous site infections, and class IV were infections of the blood-contacting surfaces or intracorporeal components of the VAD.
Results: The 8 Thoratec and 5 HeartMate patients were supported for a total of 1,648 days with a range of 33 to 279 days per patient. Every patient had at least one infection; however, there were 6 patients who had no class II or IV infections during the period of support. One of these 6 patients died of a stroke, whereas the other 5 patients survived VAD support. No trends were identified for a change in the incidence of bacterial compared with fungal infections during the course of VAD support. There was no trend for a greater number of infections in patients who died during VAD support compared with those who survived. Neither class II nor IV infections precluded transplantation. Three patients died during VAD support; 1 died as a direct consequence of fungal infection. Eight patients received transplants. One patient had an unanticipated recovery of cardiac function and the VAD was removed. Support in 1 patient is ongoing.
Conclusions: Infection during VAD support pending cardiac transplantation is an important cause of morbidity and mortality in patients maintained for longer than 30 days by circulatory assist. Infectious complications will probably be a prominent component of the risk associated with the use of chronically implanted mechanical circulatory assist devices and will likely have an important effect on the quality of life experienced by these patients.
Similar articles
-
Infections associated with ventricular assist devices: epidemiology and effect on prognosis after transplantation.Transpl Infect Dis. 2007 Jun;9(2):114-20. doi: 10.1111/j.1399-3062.2006.00185.x. Transpl Infect Dis. 2007. PMID: 17461996
-
Results of extended bridge to transplantation: window into the future of permanent ventricular assist devices.Ann Thorac Surg. 1996 Jan;61(1):396-8; discussion 407. doi: 10.1016/0003-4975(95)01020-3. Ann Thorac Surg. 1996. PMID: 8561613
-
Heart transplantation in the patient under ventricular assist complicated with device-related infection.Transplant Proc. 2004 Oct;36(8):2377-9. doi: 10.1016/j.transproceed.2004.08.113. Transplant Proc. 2004. PMID: 15561254
-
Cardiac prosthesis as an advanced surgical therapy for end-stage cardiac patients: current status and future perspectives.J Med Dent Sci. 2000 Sep;47(3):151-65. J Med Dent Sci. 2000. PMID: 12160228 Review.
-
[Clinical use of the artificial heart, indications and results].Wien Klin Wochenschr. 1988 Mar 18;100(6):161-7. Wien Klin Wochenschr. 1988. PMID: 3287767 Review. German.
Cited by
-
End-stage heart failure and mechanical circulatory support: feasibility of discharge from hospital.Neth Heart J. 2007;15(2):45-50. doi: 10.1007/BF03085953. Neth Heart J. 2007. PMID: 17612659 Free PMC article.
-
Development of a murine ventricular assist device transcutaneous drive-line model.J Heart Lung Transplant. 2008 Jul;27(7):812-4. doi: 10.1016/j.healun.2008.04.003. Epub 2008 Jun 2. J Heart Lung Transplant. 2008. PMID: 18582816 Free PMC article. No abstract available.
-
Ventricular assist devices as a bridge to cardiac transplantation. A prelude to destination therapy.Ann Surg. 1997 Jun;225(6):695-704; discussion 705-6. doi: 10.1097/00000658-199706000-00007. Ann Surg. 1997. PMID: 9230810 Free PMC article.
-
Left ventricular assist device as bridge to transplantation in patients with end-stage heart failure: Eight-year experience with the implantable HeartMate LVAS.Neth Heart J. 2002 Jun;10(6):267-271. Neth Heart J. 2002. PMID: 25696107 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous