Ex vivo evaluation of ferromagnetism, heating, and artifacts of breast tissue expanders exposed to a 1.5-T MR system
- PMID: 8574049
- DOI: 10.1002/jmri.1880050522
Ex vivo evaluation of ferromagnetism, heating, and artifacts of breast tissue expanders exposed to a 1.5-T MR system
Abstract
Three breast tissue expanders were evaluated for compatibility with MR imaging (1.5 Tesla). The metallic components of the breast tissue expanders were shown to be nonferromagnetic, heating < or = .2 degrees C and the artifacts varied. These results indicate that MR procedures may be performed safely in patients with these implants; however, artifacts may obscure implant leaks or breast lesions if located near and metal portion of the breast tissue expanders.