The efficacy of filters used in respiratory function apparatus
- PMID: 8586789
- DOI: 10.1016/0195-6701(95)90067-5
The efficacy of filters used in respiratory function apparatus
Abstract
The ability of two low resistance barrier filters (Collins DC-1 and Pall Pf 305) to remove bacteria from expired air was assessed. A specially designed coupling device was used to hold each filter or a disposable plain cardboard mouthpiece a fixed distance (4.5 cm) from a blood agar plate. Volunteers performed maximal forced vital capacity manoeuvres through the assembled apparatus and bacteria impinged on to the agar plate were enumerated. Both filters allowed the transmission of approximately one-third of expired colony forming units. The efficacy of these filters for reducing the likelihood of cross-infection during spirometry is not supported by this study.
Similar articles
-
Clinical evaluation of a screen pneumotachograph as an in-line filter.Eur Respir J. 2007 Aug;30(2):358-63. doi: 10.1183/09031936.06.00148105. Epub 2006 Jun 28. Eur Respir J. 2007. PMID: 16807260 Clinical Trial.
-
Laboratory evaluation of a filter for the control of cross-infection during pulmonary function testing.J Hosp Infect. 1992 Mar;20(3):193-8. doi: 10.1016/0195-6701(92)90087-3. J Hosp Infect. 1992. PMID: 1348774
-
Do in-line respiratory filters protect patients? Comparing bacterial removal efficiency of six filters.Pediatr Pulmonol. 2002 Nov;34(5):336-41. doi: 10.1002/ppul.10171. Pediatr Pulmonol. 2002. PMID: 12357477
-
How useful are microbial filters in respiratory apparatus?J Hosp Infect. 1997 Dec;37(4):263-72. doi: 10.1016/s0195-6701(97)90142-0. J Hosp Infect. 1997. PMID: 9457603 Review.
-
Infection control recommendations for patients with cystic fibrosis: Microbiology, important pathogens, and infection control practices to prevent patient-to-patient transmission.Am J Infect Control. 2003 May;31(3 Suppl):S1-62. Am J Infect Control. 2003. PMID: 12762292 Review. No abstract available.
Cited by
-
High-throughput molecular identification of Staphylococcus spp. isolated from a clean room facility in an environmental monitoring program.BMC Res Notes. 2010 Nov 4;3:278. doi: 10.1186/1756-0500-3-278. BMC Res Notes. 2010. PMID: 21047438 Free PMC article.
-
Joint Indian Chest Society-National College of Chest Physicians (India) guidelines for spirometry.Lung India. 2019 Apr;36(Supplement):S1-S35. doi: 10.4103/lungindia.lungindia_300_18. Lung India. 2019. PMID: 31006703 Free PMC article.
-
The practical application and interpretation of simple lung function tests in cystic fibrosis.J R Soc Med. 1999;92 Suppl 37(Suppl 37):2-12. doi: 10.1177/014107689909237S02. J R Soc Med. 1999. PMID: 10472246 Free PMC article. No abstract available.
-
Measurement of chest wall motion using a motion capture system with the one-pitch phase analysis method.Sci Rep. 2021 Nov 2;11(1):21497. doi: 10.1038/s41598-021-01033-8. Sci Rep. 2021. PMID: 34728740 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical