Impact of using an indwelling introducer on diagnosis of Swan-Ganz pulmonary artery catheter colonization
- PMID: 8641307
- DOI: 10.1007/BF01586188
Impact of using an indwelling introducer on diagnosis of Swan-Ganz pulmonary artery catheter colonization
Abstract
A prospective study was conducted to determine how the use of an indwelling introducer influences the diagnosis of pulmonary artery catheter (PAC) colonization. Sixty-six consecutive PACs and introducers were aseptically removed over a 15-month period. Two segments of the catheter and the catheter tip (a proximal segment from the portion of the catheter beneath the introducer) and two segments of the introducer (a proximal intradermal segment and the introducer tip) were cultured using a semiquantitative technique. Nineteen of 66 (28.7%) PACs showed colonization, representing an incidence of 5.6 episodes per 100 catheterization-days. Catheter tip cultures identified only 68% of colonized PACs; this yield rose to 91% when introducer tip cultures were added. These results indicate a need to evaluate both introducer tip cultures and catheter tip cultures for an accurate diagnosis of PAC colonization when an indwelling introducer is used.
Similar articles
-
Mechanisms and risk factors for infection of pulmonary artery catheters and introducer sheaths in cancer patients admitted to an intensive care unit.J Hosp Infect. 2001 Aug;48(4):289-97. doi: 10.1053/jhin.2001.1014. J Hosp Infect. 2001. PMID: 11461130
-
The pathogenesis and epidemiology of catheter-related infection with pulmonary artery Swan-Ganz catheters: a prospective study utilizing molecular subtyping.Am J Med. 1991 Sep 16;91(3B):197S-205S. doi: 10.1016/0002-9343(91)90369-9. Am J Med. 1991. PMID: 1928165 Clinical Trial.
-
Specific problems of arterial, Swan-Ganz, and hemodialysis catheters.Nutrition. 1997 Apr;13(4 Suppl):36S-41S. doi: 10.1016/s0899-9007(97)00221-9. Nutrition. 1997. PMID: 9178309 Review.
-
Influence of sterile protective sleeves on the sterility of pulmonary artery catheters.Crit Care Med. 1987 May;15(5):499-502. doi: 10.1097/00003246-198705000-00009. Crit Care Med. 1987. PMID: 3568714
-
Pulmonary artery pseudoaneurysm after Swan-Ganz catheterization: a case presentation and review of literature.Eur Heart J Acute Cardiovasc Care. 2014 Sep;3(3):281-8. doi: 10.1177/2048872613520252. Epub 2014 Jan 27. Eur Heart J Acute Cardiovasc Care. 2014. PMID: 24470440 Review.