Automated suprathreshold static perimetry screening for detecting neuro-ophthalmologic disease
Abstract
Purpose: To devise and evaluate a rapid, accurate, and cost-effective method of detecting neuro-ophthalmologic visual field defects.
Methods: One hundred fifty-nine consecutive patients were evaluated with 76-point, central 30 degree automated static threshold perimetry on the Humphrey Visual Field Analyzer, as well as by a 76-point, central 30 degree suprathreshold examination with the central reference levels set at 2 or 4 dB lower than the estimated normal median central reference level adjusted for age. Six masked readers reviewed the fields. Their readings were compared with those of the other observers, as well as with the final diagnoses as determined from all available clinical information.
Results: In detecting abnormality, the full-threshold 30 degree test had a sensitivity (percent of eyes with true field defects identified by the field test) of 93 percent or 99 percent (depending on whether borderline results were counted as a positive or negative test) and a specificity (percent of cases without true field defects appropriately identified by the field test) of 71 percent or 91 percent. In comparison, the 4-dB offset suprathreshold test had a sensitivity (averaged over all reviewers) of 79 percent or 87 percent and a specificity of 81 percent or 89 percent, whereas the 2-dB test had a sensitivity of 87 percent or 94 percent and a specificity of 73 percent or 85 percent. The mean duration of the suprathreshold tests was 3.5 +/- 1.0 minute, compared with 14.8 +/- 2.8 minutes for the full-threshold technique.
Conclusion: The central 30 degree, 76-point, 2-dB offset suprathreshold automated perimetry is more rapid and nearly as effective as the full-threshold test in detecting visual field abnormalities due to neuro-ophthalmologic disease. More quantitative, full-threshold perimetric strategies should be used in all equivocal cases and to follow progression of established disease.
Similar articles
-
Can Swedish interactive thresholding algorithm fast perimetry be used as an alternative to goldmann perimetry in neuro-ophthalmic practice?Arch Ophthalmol. 2002 Sep;120(9):1162-73. doi: 10.1001/archopht.120.9.1162. Arch Ophthalmol. 2002. PMID: 12215089
-
Short-wavelength automated perimetry in neuro-ophthalmologic disorders.Arch Ophthalmol. 1995 Apr;113(4):475-81. doi: 10.1001/archopht.1995.01100040095033. Arch Ophthalmol. 1995. PMID: 7710398
-
Sensitivity and specificity of the 76-suprathreshold visual field test to detect eyes with visual field defect by Humphrey threshold testing in a population-based setting: the Thessaloniki eye study.Am J Ophthalmol. 2004 Mar;137(3):420-5. doi: 10.1016/j.ajo.2003.09.045. Am J Ophthalmol. 2004. PMID: 15013863
-
Pitfalls of automated perimetry in glaucoma diagnosis.Curr Opin Ophthalmol. 1995 Apr;6(2):46-51. doi: 10.1097/00055735-199504000-00008. Curr Opin Ophthalmol. 1995. PMID: 10150857 Review.
-
Techniques and developments in automated perimetry: a review.Ophthalmic Physiol Opt. 1988;8(3):295-308. doi: 10.1016/0275-5408(88)90180-9. Ophthalmic Physiol Opt. 1988. PMID: 3077648 Review.
Cited by
-
Visual field loss and risk of fractures in older women.J Am Geriatr Soc. 2009 Oct;57(10):1825-32. doi: 10.1111/j.1532-5415.2009.02432.x. Epub 2009 Aug 21. J Am Geriatr Soc. 2009. PMID: 19702619 Free PMC article.
-
Use of a portable head mounted perimetry system to assess bedside visual fields.Br J Ophthalmol. 2000 Oct;84(10):1185-90. doi: 10.1136/bjo.84.10.1185. Br J Ophthalmol. 2000. PMID: 11004108 Free PMC article. Clinical Trial.
-
Sources of binocular suprathreshold visual field loss in a cohort of older women being followed for risk of falls (an American Ophthalmological Society thesis).Trans Am Ophthalmol Soc. 2007;105:312-29. Trans Am Ophthalmol Soc. 2007. PMID: 18427619 Free PMC article.
-
Mathematical modeling approaches in the study of glaucoma disparities among people of African and European descents.J Coupled Syst Multiscale Dyn. 2013 Apr 1;1(1):1-21. doi: 10.1166/jcsmd.2013.1004. J Coupled Syst Multiscale Dyn. 2013. PMID: 24501718 Free PMC article.
-
Effect of presbyopic correction lens on humphrey visual field testing in patients with multifocal intraocular lens.Medicine (Baltimore). 2023 Oct 13;102(41):e35544. doi: 10.1097/MD.0000000000035544. Medicine (Baltimore). 2023. PMID: 37832110 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous