Automated HPLC screening of newborns for sickle cell anemia and other hemoglobinopathies
- PMID: 8653895
Automated HPLC screening of newborns for sickle cell anemia and other hemoglobinopathies
Abstract
Automated HPLC is used to test dried blood-spot specimens from newborns for hemoglobins (Hb) F, A, S, C, E, and D. We present the method and report on its performance determined during >4 years of testing 2.5 x 10(6) newborns. The method features automated derivation of presumptive phenotypes; quantitative quality control and proficiency testing; throughput of one specimen per minute; small sample volume; hemoglobin concentrations quantified with an interlaboratory CV of 14-18%; retention times with interlaboratory CV of <2% and matching, within +/- 0.03 min, of laboratories and reagent lots; control of peak resolution; 0.5% detection limit for Hb S and C, and 1.0% for Hb F, A, E, and D; few interferences; and negligible background and carryover. Shortcomings of the method are the absence of microplate barcode identification and the need for manually pipetting the sample eluate into the microplate.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical