[Report of a case with re-mitral valve replacement for Björk-Shiley Delrin disc valve prosthesis--disc wear of the Delrin disc 20 years after implantation]
- PMID: 8666877
[Report of a case with re-mitral valve replacement for Björk-Shiley Delrin disc valve prosthesis--disc wear of the Delrin disc 20 years after implantation]
Abstract
We reported a patient in whom slight congestive heart failure gradually developed 20 years after mitral valve replacement with a Delrin disc Björk-Shiley valve prosthesis. Although no evident cause of prosthetic valve malfunction could be detected preoperatively, the mitral prosthesis was excised and replaced uneventfully with a 29 mm St. Jude Medical valve prosthesis. At gross inspection, marked wears of the Delrin disc surface and strut shaped indentations were present. These disc variance, occurred much earlier than initially predicted by Dr. Björk, allowed prosthetic valve malfunction. The patient with this specific model should be carefully followed-up, if necessary, performed reoperation.
Similar articles
-
[Late malfunction of the Björk-Shiley valve prosthesis due to Delrin disc defacement].Kyobu Geka. 1993 Nov;46(12):1059-62. Kyobu Geka. 1993. PMID: 8230934 Japanese.
-
[Malfunction of the Björk-Shiley valve prosthesis due to disc defacement--report of two cases of successful reoperation].Nihon Kyobu Geka Gakkai Zasshi. 1989 Mar;37(3):522-8. Nihon Kyobu Geka Gakkai Zasshi. 1989. PMID: 2768926 Japanese.
-
A case report of valve dysfunction associated with abrasion of the Delrin disk used in early Björk-Shiley mitral valves requiring resurgery.Ann Thorac Cardiovasc Surg. 2009 Apr;15(2):126-8. Ann Thorac Cardiovasc Surg. 2009. PMID: 19471217
-
[Minor strut fracture of the Björk-Shiley mitral valve].Nihon Kyobu Geka Gakkai Zasshi. 1990 Jun;38(6):1049-52. Nihon Kyobu Geka Gakkai Zasshi. 1990. PMID: 2204665 Review. Japanese.
-
[A case of acute extrinsic malfunction of prosthetic valve due to suture tag causing incomplete closure of the disc].Kyobu Geka. 1992 Jul;45(7):635-9. Kyobu Geka. 1992. PMID: 1619830 Review. Japanese.