Transmission imaging of large attenuators using a slant hole collimator on a three-headed SPECT system
- PMID: 8668108
- DOI: 10.1118/1.597715
Transmission imaging of large attenuators using a slant hole collimator on a three-headed SPECT system
Abstract
By combining conjugate views, truncation-free attenuation profiles of patients can be obtained by using slant hole collimators on three-headed SPECT systems. The alterations in reconstruction algorithms necessary for use with slant hole collimators and potential image artifacts are discussed. Based on an evaluation of the size of objects that can be imaged without truncation and the size of the overlap region in the conjugate views, a 15 degrees slant angle was determined to be optimal. Studies with a 30 degrees slant hole collimator verified the ability of slant hole transmission imaging to provide accurate, truncation-free attenuation maps of a 56 cm lateral width phantom. The center of rotation was determined to be dependent on the slant angle and radius of rotation of the slant collimator. These studies also demonstrated that the spatial resolution in the transaxial plane of the attenuation maps depends on radius of rotation of the slant hole collimator, but does not depend on the radius of rotation of an uncollimated transmission source. A multiline transmission source was investigated for use with estimating the attenuation map in Tc-99m labeled sestamibi perfusion imaging.
Similar articles
-
Estimation of attenuation maps from scatter and photopeak window single photon-emission computed tomographic images of technetium 99m-labeled sestamibi.J Nucl Cardiol. 1997 Jan-Feb;4(1 Pt 1):42-51. doi: 10.1016/s1071-3581(97)90048-9. J Nucl Cardiol. 1997. PMID: 9138839
-
Geometric Calibration and Image Reconstruction for a Segmented Slant-Hole Stationary Cardiac SPECT System.J Nucl Med Technol. 2015 Jun;43(2):103-12. doi: 10.2967/jnmt.114.153668. Epub 2015 May 8. J Nucl Med Technol. 2015. PMID: 25956691 Free PMC article.
-
Segmented slant hole collimator for stationary cardiac SPECT: Monte Carlo simulations.Med Phys. 2015 Sep;42(9):5426-34. doi: 10.1118/1.4928484. Med Phys. 2015. PMID: 26328991 Free PMC article.
-
Application of convergent-beam collimation and simultaneous transmission emission tomography to cardiac single-photon emission computed tomography.Semin Nucl Med. 1994 Jan;24(1):17-37. doi: 10.1016/s0001-2998(05)80247-8. Semin Nucl Med. 1994. PMID: 8122126 Review.
-
Attenuation compensation for cardiac single-photon emission computed tomographic imaging: Part 1. Impact of attenuation and methods of estimating attenuation maps.J Nucl Cardiol. 1995 Nov-Dec;2(6):513-24. doi: 10.1016/s1071-3581(05)80044-3. J Nucl Cardiol. 1995. PMID: 9420834 Review.
Cited by
-
Estimation of attenuation maps from scatter and photopeak window single photon-emission computed tomographic images of technetium 99m-labeled sestamibi.J Nucl Cardiol. 1997 Jan-Feb;4(1 Pt 1):42-51. doi: 10.1016/s1071-3581(97)90048-9. J Nucl Cardiol. 1997. PMID: 9138839
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources