Influence of biomaterial surface texture on bone ingrowth in the rabbit femur
- PMID: 8676259
- DOI: 10.1002/jor.1100140316
Influence of biomaterial surface texture on bone ingrowth in the rabbit femur
Abstract
The purpose of this study was to examine both the histologic and the mechanical characteristics of bone apposition to an experimental surface, arc-deposited titanium, in a rabbit model and to compare them with those of four previously studied surfaces: one layer of cobalt-chromium beads, three layers of cobalt-chromium beads, plasma-sprayed cobalt-chromium, and uncoated titanium alloy. Bilateral cylindrical implants were press-fit into the lateral femoral condyles of 70 adult New Zealand White rabbits, which were allowed unrestricted activity and then killed at 6 or 12 weeks. The distal femora were harvested, radiographed, and prepared for either mechanical or histologic evaluation. All of the implants with coated surfaces had significantly greater shear strength than the implants of grit-blasted titanium alloy after both 6 and 12 weeks. After 6 weeks, maximum bone apposition occurred in the beaded surfaces. After 12 weeks, the shear strengths and bone apposition of implants of arc-deposited titanium and of one and three layers of cobalt-chromium beads were significantly greater than those of implants of plasma-sprayed cobalt-chromium and grit-blasted titanium alloy. The histologic studies correlated with the mechanical results. After 12 weeks, the bone apposition and mechanical stability of arc-deposited titanium were similar to those of a single layer of beads. There appeared to be no advantage to multiple layers of beads, and the plasma-sprayed cobalt-chromium and grit-blasted titanium surfaces showed lower shear strength and bone apposition than the other groups.
Similar articles
-
Histological and mechanical comparison of hydroxyapatite-coated cobalt-chrome and titanium implants in the rabbit femur.J Appl Biomater. 1995 Winter;6(4):231-5. doi: 10.1002/jab.770060403. J Appl Biomater. 1995. PMID: 8589507
-
The influence of surface-blasting on the incorporation of titanium-alloy implants in a rabbit intramedullary model.J Bone Joint Surg Am. 1995 Sep;77(9):1380-95. doi: 10.2106/00004623-199509000-00015. J Bone Joint Surg Am. 1995. PMID: 7673290
-
Enhanced osseointegration of titanium implants with nanostructured surfaces: an experimental study in rabbits.Acta Biomater. 2015 Jan;11:494-502. doi: 10.1016/j.actbio.2014.10.017. Epub 2014 Oct 23. Acta Biomater. 2015. PMID: 25449926
-
Osseointegration of surface-blasted implants made of titanium alloy and cobalt-chromium alloy in a rabbit intramedullary model.J Biomed Mater Res. 1998 Oct;42(1):20-9. doi: 10.1002/(sici)1097-4636(199810)42:1<20::aid-jbm4>3.0.co;2-q. J Biomed Mater Res. 1998. PMID: 9740003
-
Hydroxyapatite-coated titanium for orthopedic implant applications.Clin Orthop Relat Res. 1988 Jul;(232):225-43. Clin Orthop Relat Res. 1988. PMID: 2838208 Review.
Cited by
-
Biological Safety Evaluation and Surface Modification of Biocompatible Ti-15Zr-4Nb Alloy.Materials (Basel). 2021 Feb 4;14(4):731. doi: 10.3390/ma14040731. Materials (Basel). 2021. PMID: 33557312 Free PMC article.
-
Increased osteoblast cell density on nanostructured PLGA-coated nanostructured titanium for orthopedic applications.Int J Nanomedicine. 2007;2(3):493-9. Int J Nanomedicine. 2007. PMID: 18019847 Free PMC article.
-
Coating of titanium with hydroxyapatite leads to decreased bone formation: A study in rabbits.Bone Joint Res. 2012 Jun 1;1(6):125-30. doi: 10.1302/2046-3758.16.2000050. Print 2012 Jun. Bone Joint Res. 2012. PMID: 23610682 Free PMC article.
-
High incidence of stem loosening in association with periprosthetic femur fractures in previously well-fixed cementless grit-blasted tapered-wedge stems.Int Orthop. 2015 Sep;39(9):1689-93. doi: 10.1007/s00264-014-2586-5. Epub 2014 Nov 11. Int Orthop. 2015. PMID: 25385003
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous