Long-term experience with the double-cuff AMS 800 artificial urinary sphincter
- PMID: 8677584
- DOI: 10.1016/S0090-4295(96)00042-8
Long-term experience with the double-cuff AMS 800 artificial urinary sphincter
Abstract
Objectives: To assess the efficacy and safety of the double-cuff artificial urinary sphincter over a long-term period.
Methods: Ninety-five patients charts were reviewed from December 1986 to November 1995. Data on the degree of urinary incontinence and complications were compiled and tabulated.
Results: There were 10 cuff erosions (10.5%) and one infection (1.1%) requiring removal of the double-cuff system, with one death from unrelated causes. Two patients in the erosion group had a double-cuff system reinserted at a later date. Eighty-six patients have a double-cuff system, with 97.6% remaining dry.
Conclusions: A tandemly placed double-cuff artificial urinary sphincter is safe and effective in the treatment of severe postprostatectomy urinary incontinence.
Similar articles
-
Long-term follow-up of single versus double cuff artificial urinary sphincter insertion for the treatment of severe postprostatectomy stress urinary incontinence.Urology. 2008 Jan;71(1):90-3. doi: 10.1016/j.urology.2007.08.017. Urology. 2008. PMID: 18242372
-
Erosion rate of the double cuff AMS800 artificial urinary sphincter: long-term followup.J Urol. 1996 Oct;156(4):1300-1. J Urol. 1996. PMID: 8808858 Clinical Trial.
-
[Artificial sphincter in severe urinary incontinence: 9 years of experience].Arch Esp Urol. 2000 Jun;53(5):409-16. Arch Esp Urol. 2000. PMID: 10961004 Spanish.
-
[Lower urinary tract reconstruction by augmentation cystoplasty and insertion of artificial urinary sphincter cuff only: long term follow-up].Prog Urol. 2004 Jun;14(3):310-4. Prog Urol. 2004. PMID: 15373171 Review. French.
-
[Results of the AMS 800 artificial urinary sphincter in men, based on a series of 47 patients].Prog Urol. 2005 Apr;15(2):244-9. Prog Urol. 2005. PMID: 15999601 Review. French.
Cited by
-
Long-term results of the implantation of the AMS 800 artificial sphincter for post-prostatectomy incontinence: a single-center experience.Int Braz J Urol. 2018 Jan-Feb;44(1):114-120. doi: 10.1590/S1677-5538.IBJU.2017.0165. Int Braz J Urol. 2018. PMID: 29211407 Free PMC article.
-
Artificial urinary sphincter: long-term results and patient satisfaction.Adv Urol. 2012;2012:835290. doi: 10.1155/2012/835290. Epub 2012 Mar 26. Adv Urol. 2012. PMID: 22536227 Free PMC article.
-
[Established treatment options for male stress urinary incontinence].Urologe A. 2007 Mar;46(3):244-8, 250-6. doi: 10.1007/s00120-007-1304-y. Urologe A. 2007. PMID: 17295032 Review. German.
-
Standing cough test stratification of moderate male stress urinary incontinence.Int Braz J Urol. 2021 Mar-Apr;47(2):415-422. doi: 10.1590/S1677-5538.IBJU.2020.0551. Int Braz J Urol. 2021. PMID: 33284545 Free PMC article.
-
Artificial urinary sphincters for male stress urinary incontinence: current perspectives.Med Devices (Auckl). 2016 Jul 4;9:175-83. doi: 10.2147/MDER.S93637. eCollection 2016. Med Devices (Auckl). 2016. PMID: 27445509 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources