Aneurysm clip testing for ferromagnetic properties: clip variability issues
- PMID: 8685361
- DOI: 10.1148/radiology.200.2.8685361
Aneurysm clip testing for ferromagnetic properties: clip variability issues
Abstract
To assess ferromagnetic properties of intracranial aneurysm clips reported to be nonferromagnetic, 1,765 Yasargil, 11 Sugita, and 15 Perneczky aneurysm clips were studied for rotation or translation on plate glass in a 1.5-T MR imager. Sixty-three clips (52 Yasargil, 11 Perneczky) weakly reoriented along the static magnetic resonance (MR) field. These results confirm the need for standardized testing for ferromagnetic properties for implantable metallic devices.
Comment in
-
Aneurysm clip testing for ferromagnetic properties: clip variability issues.Radiology. 1997 Mar;202(3):637; discussion 638-9. doi: 10.1148/radiology.202.3.9051008. Radiology. 1997. PMID: 9051008 No abstract available.
-
Aneurysm clip testing for ferromagnetic properties: clip variability issues.Radiology. 1997 Mar;202(3):640; discussion 641-2. doi: 10.1148/radiology.202.3.9051009. Radiology. 1997. PMID: 9051009 No abstract available.