Low energy internal cardioversion of atrial fibrillation resistant to transthoracic shocks
- PMID: 8697172
- PMCID: PMC484392
- DOI: 10.1136/hrt.75.6.635
Low energy internal cardioversion of atrial fibrillation resistant to transthoracic shocks
Abstract
Objective: To investigate the efficacy of internal cardioversion using low energy shocks delivered with a biatrial electrode configuration in chronic atrial fibrillation resistant to transthoracic shocks.
Methods: Low energy internal cardioversion was attempted in 11 patients who had been in atrial fibrillation for 233 (SD 193) days and had failed to cardiovert with transthoracic shocks of 360 J in both apex-base and anterior-posterior positions. Synchronised biphasic shocks of up to 400 V (approximately 6 J) were delivered, usually with intravenous sedation only, between high surface area electrodes in the right atrium and the left atrium (coronary sinus in nine, left pulmonary artery in one, left atrium via patent foramen ovale in one).
Results: Sinus rhythm was restored in 8/11 patients. The mean leading edge voltage of successful shocks was 363 (46) V [4.9 (1.2) J]. Higher energy shocks induced transient bradycardia [time to first R wave 1955 (218) ms]. No proarrhythmia or other acute complications were observed.
Conclusions: Low energy internal cardioversion of atrial fibrillation can restore sinus rhythm in patients in whom conventional transthoracic shocks have failed.
Similar articles
-
Efficacy and tolerability of transvenous low energy cardioversion of paroxysmal atrial fibrillation in humans.J Am Coll Cardiol. 1995 May;25(6):1347-53. doi: 10.1016/0735-1097(94)00555-5. J Am Coll Cardiol. 1995. PMID: 7722132 Clinical Trial.
-
Transvenous cardioversion of atrial fibrillation using low-energy shocks.J Interv Card Electrophysiol. 1997 Sep;1(2):125-9. doi: 10.1023/a:1009751031170. J Interv Card Electrophysiol. 1997. PMID: 9869960 Clinical Trial.
-
Transvenous internal cardioversion for atrial fibrillation: a randomized study on defibrillation threshold and tolerability of asymmetrical compared with symmetrical shocks.Int J Cardiol. 1999 Sep 30;71(1):63-9. doi: 10.1016/s0167-5273(99)00116-3. Int J Cardiol. 1999. PMID: 10522566 Clinical Trial.
-
Transthoracic cardioversion of atrial fibrillation and flutter: standard techniques and new advances.Am J Cardiol. 1996 Oct 17;78(8A):22-6. doi: 10.1016/s0002-9149(96)00562-0. Am J Cardiol. 1996. PMID: 8903272 Review.
-
[Atrial defibrillator].Z Kardiol. 2000;89 Suppl 3:206-14. Z Kardiol. 2000. PMID: 10810804 Review. German.
Cited by
-
Factors predicting success rate and recurrence of atrial fibrillation after first electrical cardioversion in patients with persistent atrial fibrillation.Clin Cardiol. 2001 Mar;24(3):238-44. doi: 10.1002/clc.4960240313. Clin Cardiol. 2001. PMID: 11288971 Free PMC article.
-
Atrial fibrillation begets trouble.Heart. 1997 May;77(5):393-4. doi: 10.1136/hrt.77.5.393. Heart. 1997. PMID: 9196403 Free PMC article. No abstract available.
-
Atrial fibrillation recurrence after internal cardioversion: prognostic importance of electrophysiological parameters.Heart. 2002 May;87(5):443-8. doi: 10.1136/heart.87.5.443. Heart. 2002. PMID: 11997417 Free PMC article.
-
Low-energy internal cardioversion of atrial fibrillation after failed external cardioversion: Texas Heart Institute experience and review of the literature.Tex Heart Inst J. 1999;26(2):114-9. Tex Heart Inst J. 1999. PMID: 10397433 Free PMC article. Review.
-
Rate of conversion and recurrence after sotalol treatment in patients with direct current-refractory atrial fibrillation.Clin Cardiol. 2006 Feb;29(2):56-60. doi: 10.1002/clc.4960290204. Clin Cardiol. 2006. PMID: 16506639 Free PMC article.