Instrumented implant for measuring tibiofemoral forces
- PMID: 8707796
- DOI: 10.1016/0021-9290(95)00124-7
Instrumented implant for measuring tibiofemoral forces
Abstract
Experimental measurement of loads occurring in the human knee joint will allow validation of analytical models and provide data for the design of total knee implants. A customized transducer was developed to measure the dynamic tibiofemoral force and center of pressure after total knee arthroplasty. The transducer consists of a standard tibial component to which four uniaxial load cells and an additional tibial tray have been added. The transducer was calibrated using a loading device traceable to the National Institute of Standards and Technology (NIST). The transducer was accurate to within 1% in magnitude, 0.07 mm in medial/lateral location and 0.24 m in anterior/posterior location.