Bronchodilator delivery with metered-dose inhalers in mechanically-ventilated patients
- PMID: 8730023
- DOI: 10.1183/09031936.96.09030585
Bronchodilator delivery with metered-dose inhalers in mechanically-ventilated patients
Abstract
Metered-dose inhalers (MDIs) provide several advantages over nebulizers, including ease of administration, decreased cost, reliability of dosing, and freedom from contamination. However, this method of aerosol delivery has been considered ineffective in mechanically-ventilated patients because most of the aerosol deposits in the endotracheal tube and ventilator circuit. A smaller amount of aerosol from a MDI is deposited in the lower respiratory tract of mechanically-ventilated patients than in ambulatory patients, although recent studies show that a significant bronchodilator effect can still be achieved. When employed optimally, significant bronchodilation occurs with as little as 4 puffs of a sympathomimetic aerosol. Multiple factors influence the efficacy of MDIs in mechanically-ventilated patients. The method of connecting the MDI canister to the ventilator circuit has a marked effect on aerosol delivery, and other factors include the timing of MDI actuation, ventilator mode, tidal volume, circuit humidity, and duty cycle. With a proper technique of administration, a MDI serves as an effective, convenient, and safe method for delivering bronchodilator aerosols in mechanically-ventilated patients.
Similar articles
-
Bronchodilator delivery with metered-dose inhaler during mechanical ventilation.Crit Care. 2000;4(4):227-34. doi: 10.1186/cc698. Epub 2000 Jul 11. Crit Care. 2000. PMID: 11094505 Free PMC article. Review.
-
Inhalation therapy with metered-dose inhalers and dry powder inhalers in mechanically ventilated patients.Respir Care. 2005 Oct;50(10):1331-4; discussion 1344-5. Respir Care. 2005. PMID: 16185369 Review.
-
A survey of albuterol administration practices in intubated patients in the neonatal intensive care unit.Respir Care. 2002 Jan;47(1):31-8. Respir Care. 2002. PMID: 11749685
-
Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of tidal volume.Intensive Care Med. 1999 Nov;25(11):1215-21. doi: 10.1007/s001340051049. Intensive Care Med. 1999. PMID: 10654204 Clinical Trial.
-
Inhaled salbutamol dose delivered by jet nebulizer, vibrating mesh nebulizer and metered dose inhaler with spacer during invasive mechanical ventilation.Pulm Pharmacol Ther. 2017 Aug;45:159-163. doi: 10.1016/j.pupt.2017.06.004. Epub 2017 Jun 13. Pulm Pharmacol Ther. 2017. PMID: 28627376 Clinical Trial.
Cited by
-
Nebuliser therapy in the intensive care unit.Thorax. 1997 Apr;52 Suppl 2(Suppl 2):S56-9. doi: 10.1136/thx.52.2008.s56. Thorax. 1997. PMID: 9155853 Free PMC article. Review. No abstract available.
-
Administration of dry powders during respiratory supports.Ann Transl Med. 2021 Apr;9(7):596. doi: 10.21037/atm-20-3946. Ann Transl Med. 2021. PMID: 33987294 Free PMC article. Review.
-
Mechanical ventilation of the patient with severe chronic obstructive pulmonary disease.Intensive Care Med. 1998 Sep;24(9):898-910. doi: 10.1007/s001340050688. Intensive Care Med. 1998. PMID: 9803325 Review. No abstract available.
-
Bronchodilator delivery with metered-dose inhaler during mechanical ventilation.Crit Care. 2000;4(4):227-34. doi: 10.1186/cc698. Epub 2000 Jul 11. Crit Care. 2000. PMID: 11094505 Free PMC article. Review.
-
Fundamentals of aerosol therapy in critical care.Crit Care. 2016 Oct 7;20(1):269. doi: 10.1186/s13054-016-1448-5. Crit Care. 2016. PMID: 27716346 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources