Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan
- PMID: 8733493
- PMCID: PMC1090676
- DOI: 10.1136/thx.51.4.403
Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan
Abstract
Background: Patient self-management plans for asthma use peak expiratory flow (PEF) meter readings for decisions on adjusting asthma treatment. PEF meters have been shown to be inaccurate and the effect of this inaccuracy on such treatment plans has been determined.
Methods: PEF measurements were made by 127 severe asthmatic patients at least twice a day for at least two weeks using a mini-Wright meter. The daily variation from "best" PEF and the within day PEF variability were calculated before and after correction for the meter's known inaccuracy. The effect of this data correction on the number of days when trigger points were reached for changing asthma therapy was then determined.
Results: Continuous PEF readings were available from 114 subjects with a median of 157.5 days of data per subject (range 15-489 days). Correction of the PEF data led to the number of days of satisfactory asthma control being reduced in 72% of subjects with just one subject showing an increase in satisfactory control. Data correction reduced the percentage of total days of satisfactory control from 46% to 36% of days, and increased the days requiring more inhaled steroids from 33% to 36%. The days on which a course of oral corticosteroids was required increased from 16% to 23%.
Conclusions: The accuracy of PEF meters significantly influences the interpretation of currently used asthma self-management plans. Managing asthma with the corrected PEF data would have increased the amount of treatment received by these patients since the severity of the asthma was underestimated by the raw data.
Similar articles
-
Inaccuracy of portable peak flow meters: correction is not needed. Dutch CNSLD Study Group.Acta Paediatr. 1997 Aug;86(8):888-9. doi: 10.1111/j.1651-2227.1997.tb08616.x. Acta Paediatr. 1997. PMID: 9307172
-
[Effects of nonlinear error correction of measurements obtained by peak flowmeter using the Wright Scale to assess asthma attack severity in children].Srp Arh Celok Lek. 2007 May-Jun;135(5-6):310-6. doi: 10.2298/sarh0706310s. Srp Arh Celok Lek. 2007. PMID: 17633319 Serbian.
-
Compliance with and accuracy of daily self-assessment of peak expiratory flows (PEF) in asthmatic subjects over a three month period.Eur Respir J. 1996 May;9(5):880-5. doi: 10.1183/09031936.96.09050880. Eur Respir J. 1996. PMID: 8793446
-
Incorrect use of peak flow meters: are you observing your patients?J Asthma. 2014 Aug;51(6):566-72. doi: 10.3109/02770903.2014.914218. Epub 2014 May 9. J Asthma. 2014. PMID: 24720711 Review.
-
[Comparison of different types of peak flow meter and reference values for peak expiratory flow (PEF) for healthy Japanese subjects].Nihon Rinsho. 1996 Nov;54(11):2927-32. Nihon Rinsho. 1996. PMID: 8950932 Review. Japanese.
Cited by
-
Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device.Thorax. 2003 May;58(5):411-6. doi: 10.1136/thorax.58.5.411. Thorax. 2003. PMID: 12728161 Free PMC article.
-
Is home monitoring of lung function worthwhile for children with asthma?West J Med. 2001 Nov;175(5):344-5. doi: 10.1136/ewjm.175.5.344. West J Med. 2001. PMID: 11694490 Free PMC article. No abstract available.
-
Effects of equipment and technique on peak flow measurements.BMC Pulm Med. 2006 Jun 20;6:14. doi: 10.1186/1471-2466-6-14. BMC Pulm Med. 2006. PMID: 16787543 Free PMC article.