Radial artery tonometry: moderately accurate but unpredictable technique of continuous non-invasive arterial pressure measurement
- PMID: 8785142
- DOI: 10.1093/bja/76.3.405
Radial artery tonometry: moderately accurate but unpredictable technique of continuous non-invasive arterial pressure measurement
Abstract
Radial artery tonometry provides continuous measurement of non-invasive arterial pressure (CNAP) by a sensor positioned above the radial artery. An inflatable upper arm cuff enables intermittent oscillometric calibration. CNAP was compared with invasive radial artery pressure recordings from the opposite wrist in 22 high-risk surgical patients with an inter-arm oscillometric mean arterial pressure difference < or = 10 mm Hg. Oscillometric, tonometric and invasive digital pressure values, and invasive and CNAP waveforms were obtained by the same instrument (Colin BP-508). Correlation coefficients (r) of invasive vs oscillometric values (n = 481 pairs) were 0.83, 0.90 and 0.92, and mean absolute errors of oscillometry were 7.6, 4.7, and 2.6 mm Hg for systolic, diastolic and mean arterial pressures, respectively. Correlation was poor for systolic (r = 0.80), diastolic (r = 0.77) and mean (r = 0.84) invasive vs CNAP values (n = 1375). Compared with oscillometry, mean absolute errors of 15.2, 10.9 and 9.4 mm Hg for systolic, diastolic and mean CNAP, respectively, were significantly (P < 0.001) higher. Mean prediction errors of CNAP, compared with invasive values, were -5.8 (SD 14.2) mm Hg for systolic, +7.2 (8.3) mm Hg for diastolic and +3.9 (8.8) mm Hg for mean arterial pressure. Individual patient accuracy of CNAP was assessed as good (individual prediction error < or = 5 (8) mm Hg and individual absolute error < or = 10 mm Hg) in seven patients, as acceptable (< or = 10 (12) and < or = 15 mm Hg) in 11 patients, and as inadequate in four of 22 patients. Individual accuracy of oscillometry was good or acceptable in all 22 patients. The trend in CNAP changes (difference between consecutive measurements) was sufficiently accurate during induction of anaesthesia, as only 47 (7.6%), 14 (2.3%) and 27 (4.4%) of 616 systolic, diastolic and mean CNAP values differed by more than 10 mm Hg of invasive pressure trends. We conclude that: intermittent oscillometry provides accurate arterial pressure monitoring; CNAP measurements offer a reliable trend indicator of pressure changes during induction of anaesthesia and may be considered an alternative to invasive pressure measurements, should arterial cannulation be difficult in an awake patient; and accuracy of absolute CNAP values is only moderate and unpredictable, thus radial artery tonometry should not replace invasive monitoring in high-risk patients during major surgical procedures.
Similar articles
-
Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia.Br J Anaesth. 2010 Sep;105(3):264-72. doi: 10.1093/bja/aeq143. Epub 2010 Jul 13. Br J Anaesth. 2010. PMID: 20627878
-
Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement.Br J Anaesth. 2012 Feb;108(2):202-10. doi: 10.1093/bja/aer394. Epub 2011 Dec 12. Br J Anaesth. 2012. PMID: 22171358 Clinical Trial.
-
Continuous non-invasive blood pressure monitoring by brachial artery displacement method in high-risk surgical patients.Eur J Anaesthesiol. 1995 Nov;12(6):555-63. Eur J Anaesthesiol. 1995. PMID: 8665877
-
The accuracy of radial artery applanation tonometry and intra-arterial blood pressure monitoring in critically ill patients: An evidence-based review.Nurs Crit Care. 2024 Nov;29(6):1751-1757. doi: 10.1111/nicc.13006. Epub 2023 Dec 7. Nurs Crit Care. 2024. PMID: 38059687 Review.
-
Perioperative blood pressure monitoring.Best Pract Res Clin Anaesthesiol. 2019 Jun;33(2):127-138. doi: 10.1016/j.bpa.2019.05.001. Epub 2019 May 7. Best Pract Res Clin Anaesthesiol. 2019. PMID: 31582093 Review.
Cited by
-
Evaluation of noninvasive blood pressure recording by photoplethysmography in clinical studies using angiotensin challenges.Br J Clin Pharmacol. 1999 Oct;48(4):586-93. doi: 10.1046/j.1365-2125.1999.00049.x. Br J Clin Pharmacol. 1999. PMID: 10583030 Free PMC article.
-
Blood Haemoglobin Concentration Is Directly and Independently Related with Pulse Wave Velocity, a Measure of Large Artery Stiffness.J Clin Med. 2023 Dec 11;12(24):7623. doi: 10.3390/jcm12247623. J Clin Med. 2023. PMID: 38137695 Free PMC article.
-
Subject-specific estimation of central aortic blood pressure via system identification: preliminary in-human experimental study.Med Biol Eng Comput. 2014 Oct;52(10):895-904. doi: 10.1007/s11517-014-1185-3. Epub 2014 Sep 3. Med Biol Eng Comput. 2014. PMID: 25182936
-
Non-invasive continuous blood pressure monitoring: a review of current applications.Front Med. 2013 Mar;7(1):91-101. doi: 10.1007/s11684-013-0239-5. Epub 2013 Jan 23. Front Med. 2013. PMID: 23345112 Review.
-
Differences in heart rate responses to upright posture are associated with variations in the high-frequency power of heart rate variability.Am J Physiol Heart Circ Physiol. 2024 Mar 1;326(3):H479-H489. doi: 10.1152/ajpheart.00567.2023. Epub 2023 Dec 22. Am J Physiol Heart Circ Physiol. 2024. PMID: 38133619 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources