Mechanical cardiac valve prostheses: wear characteristics and magnitudes in three bileaflet valves
- PMID: 8803764
Mechanical cardiac valve prostheses: wear characteristics and magnitudes in three bileaflet valves
Abstract
Three different bileaflet mechanical heart valves were evaluated for wear and durability characterization. The designs of the three mechanical heart valves encompass both geometrical and material interface differences. The St. Jude Medical mechanical heart valve is partially characterized by a flat-on-flat leaflet-to-orifice closing stop interface with the orifice constructed of a graphite substrate with pyrolytic carbon coating. The CarboMedicsTM mechanical heart valve is partially characterized by a flat-on-cylindrical leaflet-to-orifice closing stop interface and the orifice is constructed of solid pyrolytic carbon. The Sorin Biomedica BicarbonTM mechanical heart valve is characterized by a flat-on-flat leaflet-to-orifice closing stop interface with the leaflets constructed of pyrolytic carbon and the orifice constructed of a titanium alloy. In vitro mechanical wear analysis was performed in accordance with current FDA and ISO guidelines for accelerated life and durability testing. Comparisons revealed that the St. Jude Medical mechanical heart valve had the lowest magnitude of wear, but both the St. Jude Medical and CarboMedics mechanical heart valves proved to be very wear resistant. The Sorin Biomedica Bicarbon mechanical heart valve showed an extremely high wear rate and magnitude. The overall mechanism for material removal and wear must be fully determined to assess the long term efficacy of the mechanical heart valve.
Similar articles
-
Wear assessment in bileaflet heart valves.J Heart Valve Dis. 1996 Jun;5 Suppl 1:S133-43; discussion 144-8. J Heart Valve Dis. 1996. PMID: 8803766 Review.
-
Pitfalls and outcomes from accelerated wear testing of mechanical heart valves.J Heart Valve Dis. 1996 Jun;5 Suppl 1:S124-32; discussion 144-8. J Heart Valve Dis. 1996. PMID: 8803765
-
An initial investigation into the wear and damage within the pivots of three types of bileaflet mechanical heart valves.J Heart Valve Dis. 1996 Jun;5 Suppl 1:S111-4; discussion 144-8. J Heart Valve Dis. 1996. PMID: 8803763
-
In-vitro assessment of the wear development mechanism and stabilization of wear in the Edwards MIRA/Sorin Bicarbon mechanical heart valve orifice ring.J Heart Valve Dis. 2002 May;11(3):409-18; discussion 418. J Heart Valve Dis. 2002. PMID: 12056736
-
Structural integrity assessment of heart valve prostheses: a damage tolerance analysis of the CarboMedics Prosthetic Heart Valve.J Heart Valve Dis. 1996 Jun;5 Suppl 1:S86-96. J Heart Valve Dis. 1996. PMID: 8803760 Review.
Cited by
-
Mechanical considerations for polymeric heart valve development: Biomechanics, materials, design and manufacturing.Biomaterials. 2019 Dec;225:119493. doi: 10.1016/j.biomaterials.2019.119493. Epub 2019 Sep 17. Biomaterials. 2019. PMID: 31569017 Free PMC article. Review.
-
Pre- and post-surgical evaluation of the inflammatory response in patients with aortic stenosis treated with different types of prosthesis.BMC Cardiovasc Disord. 2017 Apr 14;17(1):100. doi: 10.1186/s12872-017-0526-1. BMC Cardiovasc Disord. 2017. PMID: 28410571 Free PMC article.