In vivo accuracy testing and clinical experience with the ISG Viewing Wand
- PMID: 8805161
- DOI: 10.1097/00006123-199607000-00048
In vivo accuracy testing and clinical experience with the ISG Viewing Wand
Abstract
Objective: A frameless stereotactic system (the Viewing Wand; Elekta Instruments, Atlanta, GA) has been developed for use with preoperative computed tomography, magnetic resonance imaging, and positron emission tomography as an adjunct for surgical planning and intraoperative navigation. This clinical study was designed to evaluate the safety, efficacy, and accuracy of the Viewing Wand in a variety of intracranial procedures.
Methods: We used this system in 250 patients undergoing a wide range of neurosurgical procedures from July 1990 to July 1994, to assess its clinical usefulness and safety. In a subset of 45 neurosurgical patients studied between March 1993 and March 1994, a battery of objective accuracy measurements was obtained before and during surgery.
Results: In this series, there were no instances of adverse outcomes attributable to the use of this system. A comparison of two alternative patient-image registration techniques established that the fiducial-fit method was slightly more accurate than the surface-fit method (geometric means = 2.51 and 3.03 mm, respectively). The clinical accuracy achieved with magnetic resonance imaging was nearly equivalent to that with computed tomography.
Conclusions: On the basis of this clinical series, recommendations are made regarding preoperative scanning parameters, registration techniques, and methods for reestablishing registration if needed during the course of surgery. The primary clinical benefits of the wand in this series were improved intraoperative navigation and surgical safety. For most cases, the wand was also useful in planning the location and size of the scalp incision, craniotomy, or corticotomy, as well as the extent of surgical resection.
Similar articles
-
Clinical use of a frameless stereotactic arm: results of 325 cases.J Neurosurg. 1995 Aug;83(2):197-205. doi: 10.3171/jns.1995.83.2.0197. J Neurosurg. 1995. PMID: 7616261
-
The impact of interactive image guided surgery: the Bristol experience with the ISG/Elekta viewing Wand.Acta Neurochir Suppl. 1995;64:54-8. doi: 10.1007/978-3-7091-9419-5_12. Acta Neurochir Suppl. 1995. PMID: 8748584
-
Frameless stereotaxy and interactive neurosurgery with the ISG viewing wand.Acta Neurochir Suppl. 1995;64:49-53. doi: 10.1007/978-3-7091-9419-5_11. Acta Neurochir Suppl. 1995. PMID: 8748583
-
Image-guided surgery of epilepsy.Neurosurg Clin N Am. 1996 Apr;7(2):229-43. Neurosurg Clin N Am. 1996. PMID: 8726438 Review.
-
Minimally invasive brain surgery.Ann Acad Med Singap. 1993 May;22(3 Suppl):459-63. Ann Acad Med Singap. 1993. PMID: 8215201 Review.
Cited by
-
Computer-aided navigation in neurosurgery.Neurosurg Rev. 2003 May;26(2):73-99; discussion 100-1. doi: 10.1007/s10143-003-0262-0. Neurosurg Rev. 2003. PMID: 12962294 Review.
-
Multimodal navigated skull base tumor resection using image-based vascular and cranial nerve segmentation: A prospective pilot study.Surg Neurol Int. 2015 Nov 19;6:172. doi: 10.4103/2152-7806.170023. eCollection 2015. Surg Neurol Int. 2015. PMID: 26674155 Free PMC article.
-
Frame-based stereotactic biopsy remains an important diagnostic tool with distinct advantages over frameless stereotactic biopsy.J Neurooncol. 2005 Jun;73(2):173-9. doi: 10.1007/s11060-004-4208-3. J Neurooncol. 2005. PMID: 15981109
-
Image integration in electroanatomic mapping.Herzschrittmacherther Elektrophysiol. 2007 Sep;18(3):122-30. doi: 10.1007/s00399-007-0571-z. Herzschrittmacherther Elektrophysiol. 2007. PMID: 17891488 Review.
-
Pre-operative planning and intra-operative guidance in modern neurosurgery: a review of 300 cases.Ann R Coll Surg Engl. 1999 Jul;81(4):217-25. Ann R Coll Surg Engl. 1999. PMID: 10615186 Free PMC article. Clinical Trial.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical