The Elag-Koln automatic blood pressure recorder. A clinical appraisal
- PMID: 884029
- PMCID: PMC483318
- DOI: 10.1136/hrt.39.7.795
The Elag-Koln automatic blood pressure recorder. A clinical appraisal
Abstract
The performance of an Elag-Koln semiautomatic blood pressure recorder was compared with the London School of Hygiene mercury sphygmomanometer in 24 subjects providing a wide range of blood pressure measurements. Readings with the two instruments correlated highly (for systole, r=0-99; for diastole phase 4, r=0-97; for diastole phase 5,r=0-98), and the slopes of the regressions did not differ significantly from unity. Elag-Koln measurements were higher for systole (mean difference 3-7 mmHg,P less than 0-001) and diastole phase 5 (mean difference 7-4 mmHg,P less than 0-001), but agreed closely with diastole phase 4 readings with the London School of Hygiene instrument. The Elag-Koln recorder tested was compact, easy to use, and had acceptable accuracy. This type of instrument deserves further testing to examine its suitability for general use.
Similar articles
-
Miniature electronic blood pressure monitor compared with a blind-reading mercury sphygmomanometer in pregnancy.Eur J Obstet Gynecol Reprod Biol. 1989 Nov;33(2):147-53. doi: 10.1016/0028-2243(89)90207-4. Eur J Obstet Gynecol Reprod Biol. 1989. PMID: 2583339
-
Clinical evaluation of Dinamap 845 automated blood pressure recorder.Br Heart J. 1980 Feb;43(2):202-5. doi: 10.1136/hrt.43.2.202. Br Heart J. 1980. PMID: 7362713 Free PMC article.
-
A comparative study of automated blood pressure recorders.Postgrad Med J. 1980 Dec;56(662):815-7. doi: 10.1136/pgmj.56.662.815. Postgrad Med J. 1980. PMID: 7267490 Free PMC article. Clinical Trial.
-
[Are automatic devices suitable for blood pressure determinations?].Schweiz Med Wochenschr. 1979 Sep 1;109(33):1225-30. Schweiz Med Wochenschr. 1979. PMID: 493912 German.
-
Accuracy of the London School of Hygiene and Remler M2000 sphygmomanometers.Clin Sci (Lond). 1981 Dec;61 Suppl 7:399s-401s. doi: 10.1042/cs061399s. Clin Sci (Lond). 1981. PMID: 7318344 Clinical Trial.
Cited by
-
Nurse- and automatic machine-measured blood pressure readings: a comparative study.Public Health Rep. 1980 Jul-Aug;95(4):382-5. Public Health Rep. 1980. PMID: 7422805 Free PMC article. No abstract available.
-
The Remler M2000 semi-automatic blood pressure recorder.Br Heart J. 1979 Sep;42(3):366. doi: 10.1136/hrt.42.3.366. Br Heart J. 1979. PMID: 508463 Free PMC article. No abstract available.
-
The methodology of blood pressure recording.Br J Clin Pharmacol. 1978 Sep;6(3):193-201. doi: 10.1111/j.1365-2125.1978.tb04585.x. Br J Clin Pharmacol. 1978. PMID: 687498 Free PMC article. No abstract available.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources